Folke M, Hök B
Department of Computer Science and Electronics, Mälardalen University, Västerås, Sweden.
Med Biol Eng Comput. 2008 Jan;46(1):55-9. doi: 10.1007/s11517-007-0228-4. Epub 2007 Sep 11.
End tidal carbon dioxide measurements with an electro acoustic capnograph prototype have been demonstrated. The aim of this study was to verify that it is possible to obtain an adequate capnogram using the prototype and to investigate the influence of ambient temperature and humidity variations. By simultaneous measurements with a reference capnograph, on subjects performing exercise, hypo- and hyperventilation, P(ET)CO(2) readings from the reference were compared with the output signal from the prototype. The capnogram from the prototype correlated well with the reference in terms of breath time. The first parts of the expiration and inspiration phases were steeper for the reference than the prototype. The output signal from the prototype correlated well with the reference P(ET)CO(2) readings with a correlation coefficient of 0.93 at varied temperature and relative humidity.