Kim Hyunook, Lim Honglae, Colosimo Mark F
The University of Seoul Department of Environmental Engineering, Seoul, Republic of Korea.
J Environ Sci Health A Tox Hazard Subst Environ Eng. 2007 Sep;42(11):1665-70. doi: 10.1080/10934520701518190.
A new method for determining wastewater chemical oxygen demand (COD) using ultrasonic digestion and titration based on oxidation reduction potential (ORP) was developed. COD values of potassium hydrogen phthalate solution obtained by ultrasonic digestion were well matched with those obtained using Standard Methods. When applied to determine COD of real wastewater collected from a local treatment plant, results from the new method were within 80% to 90% of those obtained using Standard Methods. Nonetheless, the proposed strategy has the potential to be implemented into an online COD analyzing system.
开发了一种基于氧化还原电位(ORP),采用超声消解和滴定法测定废水化学需氧量(COD)的新方法。通过超声消解得到的邻苯二甲酸氢钾溶液的COD值与采用标准方法得到的值非常匹配。当应用于测定从当地污水处理厂收集的实际废水的COD时,新方法的结果在采用标准方法所得结果的80%至90%范围内。尽管如此,所提出的策略仍有潜力应用于在线COD分析系统。