Mayrovitz H N
College of Medical Sciences, Nova Southeastern University, Ft. Lauderdale, Florida 33328, USA.
Lymphology. 2007 Jun;40(2):87-94.
Overall limb lymphedema can be assessed by several methods but none are suitable to determine local edema. Quantifying local edema could provide important information not previously available. Our goal was to determine the suitability of using the tissue dielectric constant (TDC) as and index of local tissue water to detect and quantify edema in postmastectomy patients with unilateral arm lymphedema. Segmental arm volume and TDC were measured in both arms of 18 women with unilateral lymphedema, and in 15 premenopausal and 15 postmenopausal controls. TDC was measured at a frequency of 300 MHz using open-ended coaxial probes with effective measuring depths of 0.5, 1.5, 2.5 and 5.0 mm. For patients and controls, absolute TDC depended on measurement depth but for any depth the TDC of lymphedematous segments was significantly greater than for non-affected contralateral arms (p<0.001). At a depth of 2.5 mm, the TDC ratio between arms for patients was 1.64+/-0.30 vs.1.04+/-0.04 for both control groups (p<0.001). No patient's TDC ratio was as low as 1.2 and no control subject's TDC ratio was as great as 1.2. Results suggest that this method is a good quantitative discriminator of the presence of lymphedema in patients with unilateral limb lymphedema.
整体肢体淋巴水肿可通过多种方法进行评估,但没有一种方法适用于确定局部水肿情况。量化局部水肿能够提供此前无法获得的重要信息。我们的目标是确定使用组织介电常数(TDC)作为局部组织水分指标来检测和量化单侧手臂淋巴水肿的乳房切除术后患者水肿情况的适用性。对18名单侧淋巴水肿女性患者的双臂以及15名绝经前和15名绝经后对照者的双臂进行了节段性手臂体积和TDC测量。使用有效测量深度为0.5、1.5、2.5和5.0毫米的开放式同轴探头在300兆赫频率下测量TDC。对于患者和对照者,绝对TDC取决于测量深度,但在任何深度下,淋巴水肿节段的TDC均显著高于未受影响的对侧手臂(p<0.001)。在2.5毫米深度处,患者双臂之间的TDC比值为1.64±0.30,而两个对照组的该比值为1.04±0.04(p<0.001)。没有患者的TDC比值低至1.2,也没有对照者的TDC比值高达1.2。结果表明,该方法是单侧肢体淋巴水肿患者淋巴水肿存在情况的良好定量判别方法。