Korakianitis Theodosios, Shi Yubing
Department of Engineering, Queen Mary University of London, London, United Kingdom.
ASAIO J. 2007 Sep-Oct;53(5):537-48. doi: 10.1097/MAT.0b013e318142bfce.
We report the first attempt to study with numerical methods ventricular assist device (VAD) models and the effects of various inlet VAD cannulations, coupling physical explanations and numerical investigation conclusions with clinical research results. We compared the hemodynamic response with VAD support by using two distinct VAD-inlet cannulation configurations: left atrium to aorta and left ventricular apex to aorta. Impeller pump and displacement pump VADs are considered. Constant VAD flow rate and counterpulsation motion models are simulated. The native cardiovascular system is modeled using the concentrated-parameter method by considering the flow resistance, vessel elasticity, and inertial effect of blood flow in cardiovascular system individual segments. Impeller and displacement pump dynamic models are represented by corresponding inlet and outlet flow rate changes in the VADs. Results show that the two VAD inlet cannulation configurations produce similar cardiac response (flows, pressures, volumes), except that when the VAD flow approaches the 100% assisting condition, the peak left ventricular systolic pressure and diastolic volume increase slightly in the left atrial cannulation, whereas they drop markedly in the left ventricular apex cannulation, suggesting increased ventricular wall tension and ventricular dilatation in the left atrial cannulation and that hemodynamically the left ventricular apex cannulation is more advantageous.
我们报告了首次尝试使用数值方法研究心室辅助装置(VAD)模型以及各种VAD入口插管的影响,将物理解释和数值研究结论与临床研究结果相结合。我们通过使用两种不同的VAD入口插管配置比较了VAD支持下的血流动力学反应:左心房至主动脉和左心室心尖至主动脉。考虑了叶轮泵和容积泵VAD。模拟了恒定的VAD流速和反搏运动模型。通过考虑心血管系统各个节段中血流的流动阻力、血管弹性和惯性效应,使用集中参数法对天然心血管系统进行建模。叶轮泵和容积泵的动态模型由VAD中相应的入口和出口流速变化表示。结果表明,两种VAD入口插管配置产生相似的心脏反应(流量、压力、容积),只是当VAD流量接近100%辅助状态时,左心房插管时左心室收缩压峰值和舒张容积略有增加,而在左心室心尖插管时则明显下降,这表明左心房插管时心室壁张力增加和心室扩张,并且从血流动力学角度来看,左心室心尖插管更具优势。