Verberne Hein J, Habraken Jan B A, van Eck-Smit Berthe L F, Agostini Denis, Jacobson Arnold F
Department of Nuclear Medicine, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands.
Eur J Nucl Med Mol Imaging. 2008 Mar;35(3):547-53. doi: 10.1007/s00259-007-0611-2. Epub 2007 Oct 6.
There is lack of validation and standardisation of acquisition parameters for myocardial (123)I-metaiodobenzylguanidine (MIBG). This lack of standardisation hampers large scale implementation of (123)I-MIBG parameters in the evaluation of patients with chronic heart failure (CHF).
In a retrospective multi-centre study (123)I-MIBG planar scintigrams obtained on 290 CHF patients (82% male; 58% dilated cardiomyopathy; New York Heart Association [NYHA classification] > I) were reanalysed to determine the late heart-to-mediastinum ratio (H/M).
There was a large variation in acquisition parameters. Multivariate forward stepwise regression showed that a significant proportion (31%, p < 0.001) of the variation in late H/M could be explained by a model containing patient-related variables and acquisition parameters. Left ventricular ejection fraction (p < 0.001), type of collimation (p < 0.001), acquisition duration (p = 0.001), NYHA class (p = 0.028) and age (p = 0.034) were independent predictors of late H/M.
Acquisitions parameters are independent contributors to the variation of semi-quantitative measurements of cardiac (123)I-MIBG uptake. Improved standardisation of cardiac (123)I-MIBG imaging parameters would contribute to increased clinical applicability for this procedure.