Brumberg Joachim, Blazhenets Ganna, Schröter Nils, Frings Lars, Jost Wolfgang H, Lapa Constantin, Meyer Philipp T
Department of Nuclear Medicine, University Hospital Würzburg and Julius-Maximilians-University, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Department of Nuclear Medicine, Medical Center, Faculty of Medicine, University of Freiburg, Hugstetter Straße 55, 79106, Freiburg im Breisgau, Germany.
EJNMMI Phys. 2019 Jul 23;6(1):12. doi: 10.1186/s40658-019-0250-2.
The heart-to-mediastinum (H/M) ratio is a commonly used parameter to measure cardiac I-123 metaiodobenzylguanidine (MIBG) uptake. Since the H/M ratio is substantially influenced by the collimator type, we investigated whether an empirical linear conversion of H/M ratios between camera systems with low-energy (LE) and medium-energy (ME) collimator is possible.
We included 18 patients with parkinsonism who were referred to one of the two participating molecular imaging facilities for the evaluation of cardiac sympathetic innervation by MIBG scintigraphy. Two consecutive planar image datasets were acquired with LE and ME collimators at 4 h after MIBG administration. Linear regression analyses were performed to describe the association between the H/M ratios gained with both collimator settings, and the accuracy of a linear transfer of the H/M ratio between collimators and across centers was assessed using a leave-one-out procedure.
H/M ratios acquired with LE and ME collimators showed a strong linear relationship both within each imaging facility (R = 0.99, p < 0.001 and R = 0.90, p < 0.001) and across centers (H/M-LE = 0.41 × H/M-ME + 0.63, R = 0.97, p < 0.001). A linear conversion of H/M ratios between collimators and across centers was estimated to be very accurate (mean absolute error 0.05 ± 0.04; mean relative absolute error 3.2 ± 2.6%).
The present study demonstrates that a simple linear conversion of H/M ratios acquired with different collimators is possible with high accuracy. This should greatly facilitate the exchange of normative data between settings and pooling of data from different institutions.
心纵隔(H/M)比值是用于测量心脏I-123间碘苄胍(MIBG)摄取的常用参数。由于H/M比值受准直器类型的影响很大,我们研究了低能(LE)和中能(ME)准直器的相机系统之间H/M比值的经验线性转换是否可行。
我们纳入了18例帕金森病患者,他们被转诊至两个参与研究的分子影像机构之一,通过MIBG闪烁显像评估心脏交感神经支配情况。在注射MIBG后4小时,使用LE和ME准直器连续采集两个平面图像数据集。进行线性回归分析以描述两种准直器设置下获得的H/M比值之间的关联,并使用留一法评估准直器之间以及不同中心之间H/M比值线性转换的准确性。
在每个影像机构内部,LE和ME准直器获得的H/M比值均显示出很强的线性关系(R = 0.99,p < 0.001和R = 0.90,p < 0.001),并且在不同中心之间也存在线性关系(H/M-LE = 0.41 × H/M-ME + 0.63,R = 0.97,p < 0.001)。估计准直器之间以及不同中心之间H/M比值的线性转换非常准确(平均绝对误差0.05 ± 0.04;平均相对绝对误差3.2 ± 2.6%)。
本研究表明,不同准直器获得的H/M比值可以进行简单的高精度线性转换。这将极大地促进不同设备之间规范数据的交换以及不同机构数据的汇总。