Hou Keyong, Wang Junde, Li Haiyang
Dalian Institute of Chemical Physics, Chinese Academy of Sciences, 457 Zhongshan Road, Dalian 116023, P.R. China.
Rapid Commun Mass Spectrom. 2007;21(22):3554-60. doi: 10.1002/rcm.3250.
A novel membrane inlet interface coupled to a single-photon ionization (SPI) miniature time-of-flight mass spectrometer has been developed for on-line rapid measurement of volatile organic compounds (VOCs). The vacuum ultraviolet (VUV) light source for SPI was a commercial krypton discharge lamp with photon energy of 10.6 eV and photon flux of 10(10) photons/s. The experimental results showed that the sensitivity was 5 times as high as obtained with the traditional membrane inlet. The enrichment efficiency could be adjusted in the range of 10 to 20 times for different VOCs when a buffer cell was added to the inlet interface, and the memory effect was effectively eliminated. A detection limit as low as 25 parts-per-billion by volume (ppbv) for benzene has been achieved, with a linear dynamic range of three orders of magnitude. The rise times were 6 s, 10 s and 15 s for benzene, toluene and p-xylene, respectively, and the fall time was only 6 s for all of these compounds. The analytical capacity of this system was demonstrated by the on-line analysis of VOCs in single puff mainstream cigarette smoke, in which more than 50 compounds were detected in 2 s.
一种与单光子电离(SPI)微型飞行时间质谱仪联用的新型膜进样接口已被开发出来,用于在线快速测量挥发性有机化合物(VOCs)。用于SPI的真空紫外(VUV)光源是一种商业氪放电灯,光子能量为10.6 eV,光子通量为10(10) 光子/秒。实验结果表明,其灵敏度是传统膜进样方式的5倍。当在进样接口中添加一个缓冲池时,对于不同的VOCs,富集效率可在10至20倍的范围内调节,并且有效消除了记忆效应。已实现对苯的低至25体积十亿分之一(ppbv)的检测限,线性动态范围为三个数量级。苯、甲苯和对二甲苯的上升时间分别为6秒、10秒和15秒,而所有这些化合物的下降时间仅为6秒。通过对单口主流卷烟烟雾中的VOCs进行在线分析,证明了该系统的分析能力,在2秒内检测到了50多种化合物。