Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian 116023, China.
Analyst. 2013 Oct 7;138(19):5826-31. doi: 10.1039/c3an00659j.
An in-source, stretched, hollow fiber membrane (HFM) inlet has been developed to improve the sensitivity of on-line time-of-flight mass spectrometry (TOFMS) with a vacuum ultraviolet (VUV) lamp based single photon ionization (SPI) source for the direct analysis of liquid samples. A 2-cm HFM was stretched to 8 cm in length, and placed in the ion source and directly under the VUV lamp window with a distance of 15 mm. Compared with the conventional flow-through configuration under the same experimental conditions, the signal intensities of selected volatile organic compounds (VOCs) of methyl tert-butyl ether (MTBE), ethyl tert-butyl ether (ETBE), benzene, toluene and p-xylene were increased over 5-fold in magnitude, and the response time was shortened to one-third. The limits of detection (LOD) of MTBE, ETBE, benzene, toluene and p-xylene ranged from 0.25 to 1.3 μg L(-1) with a measurement time of 60 s, and three orders of linear range were obtained with correlation coefficients of 0.9972-0.9992. The present results suggest that the in-source stretched HFM is a simple and effective way to increase the sensitivity and shorten response time of the membrane inlet, and we believe that it will also be beneficial to other types of on-line mass spectrometer for the on-line analysis of VOCs in water with a VUV lamp based SPI ion source.
一种源内拉伸中空纤维膜(HFM)入口已被开发出来,以提高在线飞行时间质谱(TOFMS)的灵敏度,该质谱采用真空紫外(VUV)灯单光子电离(SPI)源,用于直接分析液体样品。将 2 厘米长的 HFM 拉伸至 8 厘米长,并将其放置在离子源中,直接置于 VUV 灯窗口下方,距离为 15 毫米。与相同实验条件下的传统流通式配置相比,选择的挥发性有机化合物(VOC)的甲基叔丁基醚(MTBE)、乙基叔丁基醚(ETBE)、苯、甲苯和对二甲苯的信号强度增加了 5 倍以上,响应时间缩短至三分之一。MTBE、ETBE、苯、甲苯和对二甲苯的检出限(LOD)范围为 0.25-1.3 μg/L,测量时间为 60 s,获得了三个数量级的线性范围,相关系数为 0.9972-0.9992。目前的结果表明,源内拉伸 HFM 是一种简单有效的方法,可以提高膜入口的灵敏度并缩短响应时间,我们相信它也将有益于其他类型的在线质谱仪,用于在线分析基于 VUV 灯 SPI 离子源的水中的 VOC。