Lane Jason W, Rehak Nadja N, Hortin Glen L, Zaoutis Theoklis, Krause Philip R, Walsh Thomas J
Microbiology Service, Department of Laboratory Medicine, W. G. Magnuson Clinical Center, NIH, Bethesda, Maryland, USA.
Clin Chim Acta. 2008 Jan;387(1-2):145-9. doi: 10.1016/j.cca.2007.08.008. Epub 2007 Aug 30.
Acute increases in serum inorganic phosphorus (Pi) up to 4.75 mmol/l in the absence of hypocalcemia and tissue deposition of calcium phosphate were noted in 3 patients receiving liposomal amphotericin B (L-AMB). We investigated L-AMB as a possible cause of pseudohyperphosphatemia.
Serum samples from the index patient were analyzed for Pi content by our laboratory's primary analyzer (Synchron LX20) and by an alternate analyzer (Vitros). Clear and lipemic serum pools, and normal saline, were spiked with L-AMB and analyzed by the LX20 Pi method. Ultrafiltration studies were performed on patient and spiked sera.
Increased Pi values were obtained only from the LX20 analyzer. There was a direct linear relationship between the concentration of L-AMB in the spiked samples and the LX20 Pi results, indicating a 0.9 mmol/l Pi increase for every 100 mg/l increase in L-AMB. Ultrafiltration normalized the Pi results.
Serum Pi results may be falsely increased in patients receiving L-AMB when measured by the LX20 analyzer. This novel cause of pseudohyperphosphatemia is due to interference of L-AMB with the method and is corrected by ultrafiltration of the specimen. Since the LX20 analyzer is widely used by the clinical laboratories clinicians and laboratory personnel should recognize this interference in order to avoid unnecessary diagnostic procedures and interventions.
在3例接受脂质体两性霉素B(L-AMB)治疗的患者中,观察到血清无机磷(Pi)急性升高至4.75 mmol/l,且无低钙血症和磷酸钙组织沉积。我们研究了L-AMB作为假性高磷血症可能病因的情况。
用本实验室的主要分析仪(Synchron LX20)和另一台分析仪(Vitros)对索引患者的血清样本进行Pi含量分析。向清亮和脂血血清池以及生理盐水中加入L-AMB,并用LX20 Pi法进行分析。对患者血清和加样血清进行超滤研究。
仅从LX20分析仪获得了升高的Pi值。加样样本中L-AMB的浓度与LX20 Pi结果之间存在直接线性关系,表明L-AMB每增加100 mg/l,Pi增加0.9 mmol/l。超滤使Pi结果正常化。
使用LX20分析仪测量时,接受L-AMB治疗的患者血清Pi结果可能会被错误升高。这种假性高磷血症的新病因是L-AMB对该方法的干扰,通过对样本进行超滤可纠正。由于临床实验室广泛使用LX20分析仪,临床医生和实验室人员应认识到这种干扰,以避免不必要的诊断程序和干预措施。