The aim of this study was to evaluate the reliability of isokinetic and isometric assessments of the knee extensor and the flexor muscle function using the Con-Trex isokinetic dynamometer. Thirty healthy subjects (15 males, 15 females) were tested and retested 7 days later for maximal strength (isokinetic peak torque, work, power and angle of peak torque as well as isometric maximal voluntary contraction torque and rate of torque development) and fatigue (per cent loss and linear slope of torque and work across a series of 20 contractions). For both the knee extensor and the flexor muscle groups, all strength data - except angle of peak torque - demonstrated moderate-to-high reliability, with intraclass correlation coefficients (ICC) higher than 0.86. The highest reliability was observed for concentric peak torque of the knee extensor muscles (ICC = 0.99). Test-retest reliability of fatigue variables was moderate for the knee extensor (ICC range 0.84-0.89) and insufficient-to-moderate for the knee flexor muscles (ICC range 0.78-0.81). The more reliable index of muscle fatigue was the linear slope of the decline in work output. These findings establish the reliability of isokinetic and isometric measurements using the Con-Trex machine.
本研究旨在使用Con-Trex等速测力计评估膝关节伸肌和屈肌功能的等速和等长评估的可靠性。对30名健康受试者(15名男性,15名女性)进行了测试,并在7天后重新测试了最大力量(等速峰值扭矩、功、功率和峰值扭矩角度以及等长最大自主收缩扭矩和扭矩发展速率)和疲劳(在一系列20次收缩过程中扭矩和功的损失百分比和线性斜率)。对于膝关节伸肌和屈肌组,除峰值扭矩角度外,所有力量数据均显示出中到高的可靠性,组内相关系数(ICC)高于0.86。膝关节伸肌的同心峰值扭矩观察到最高的可靠性(ICC = 0.99)。膝关节伸肌疲劳变量的重测可靠性中等(ICC范围0.84 - 0.89),膝关节屈肌的可靠性不足到中等(ICC范围0.78 - 0.81)。肌肉疲劳更可靠的指标是功输出下降的线性斜率。这些发现确立了使用Con-Trex机器进行等速和等长测量的可靠性。