Yeoh Wee-Soon, Zhang Cishen, Chen Ming
Sch. of Electr. & Electron. Eng., Nanyang Technol. Univ., Singapore.
Conf Proc IEEE Eng Med Biol Soc. 2006;2006:2758-61. doi: 10.1109/IEMBS.2006.260228.
This paper describes how the recently developed constrained least squares (CLS) filtering algorithm can be made iterative to improve the resolution gain (RG) of medical ultrasound images. We propose the use of the iterative CLS (ICLS) filter, by incorporating the recently proposed ultrasound tissue model, to account for the random fluctuations of the tissue signal within the received ultrasound radio frequency (RF) echo signal. The resulting improvement in RG is demonstrated by eight different abdomen ultrasound images where progressive improvements in both the axial and lateral directions can be observed.
本文描述了如何使最近开发的约束最小二乘(CLS)滤波算法具有迭代性,以提高医学超声图像的分辨率增益(RG)。我们建议使用迭代CLS(ICLS)滤波器,通过纳入最近提出的超声组织模型,来考虑接收到的超声射频(RF)回波信号中组织信号的随机波动。通过八幅不同的腹部超声图像证明了RG的改善,在这些图像中可以观察到轴向和横向方向都有逐步改善。