System miniaturization and steady progress towards a totally implantable prosthetic system is the current trend in cochlear implant technology. To achieve this objective, the external microphone of present implants needs to be implantable. This goal can be accomplished by placing a miniature accelerometer on the ossicular chain in the middle ear to detect and convert bone vibrations into an electrical signal for further processing and stimulating cochlear electrodes. This paper describes the characterization of the umbo of a human temporal bone before and after the removal of the incus to determine the impact of the resulting change in umbo mechanics and attached accelerometer performance. With the removal of the incus, the umbo vibration acceleration frequency response in the direction perpendicular to the tympanic membrane increases by 5 dB below 2 kHz. Above 2 kHz the response diverges due to the change of ossicular chain resonant frequency caused by the removal of the incus. However, at each frequency the umbo vibration acceleration exhibits a linear function of the input sound pressure level (SPL) with a slope of 20 dB per decade before and after removal of the incus. A commercial accelerometer attached to the umbo shows similar characteristics. From the measurement results of umbo characterization, a miniaturized implantable accelerometer with a packaged mass below 20 milligrams, a sensing resolution of 35 microg rms/square root Hz, and a bandwidth of 10 kHz would be required to detect normal conversation.
系统小型化以及朝着完全可植入式假体系统稳步发展是当前人工耳蜗技术的趋势。为实现这一目标,目前植入设备的外部麦克风需要能够植入。通过在中耳的听骨链上放置一个微型加速度计来检测骨振动并将其转换为电信号以进行进一步处理和刺激耳蜗电极,这一目标便可达成。本文描述了在切除砧骨前后对人颞骨鼓膜脐部的特性进行研究,以确定砧骨切除后鼓膜脐部力学变化及所附着加速度计性能的影响。切除砧骨后,在垂直于鼓膜的方向上,鼓膜脐部振动加速度频率响应在2 kHz以下增加了5 dB。高于2 kHz时,由于切除砧骨导致听骨链共振频率发生变化,响应出现偏差。然而,在每个频率下,切除砧骨前后鼓膜脐部振动加速度均呈现出输入声压级(SPL)的线性函数关系,斜率为每十倍频程20 dB。附着在鼓膜脐部的商用加速度计也显示出类似的特性。根据鼓膜脐部特性的测量结果,若要检测正常对话,需要一个封装质量低于20毫克、传感分辨率为35 μg rms/√Hz且带宽为10 kHz的小型化可植入加速度计。