用于帕金森病丘脑底核刺激的Itrel II脉冲发生器的使用寿命。

Lifetime of Itrel II pulse generators for subthalamic nucleus stimulation in Parkinson's disease.

作者信息

Anheim Mathieu, Fraix Valérie, Chabardès Stéphan, Krack Paul, Benabid Alim-Louis, Pollak Pierre

机构信息

Department of Neurology, University Hospital A. Michallon, Grenoble, France.

出版信息

Mov Disord. 2007 Dec;22(16):2436-9. doi: 10.1002/mds.21726.

Abstract

The efficacy of bilateral subthalamic nucleus (STN) stimulation in Parkinson's disease (PD) is well-established but little is known about the lifetime of implanted pulse generators (IPG). To investigate the lifetime of the bilaterally implanted Itrel II(R) (Medtronic, Minneapolis) pulse generator, the first 49 consecutive patients with PD having been operated on at our center for bilateral STN chronic stimulation were reviewed with noting of the stimulation parameters in use prior to IPG replacement. The mean electrical voltage was 3.2 +/- 0.3 V, mean pulse width was 65 +/- 10 mus, and mean frequency was 145 +/- 16 Hz. Replacement of an IPG was anticipated in 25% due to unilateral low-battery signaling, or end of life. In either case, replacement of the contralateral IPG was undertaken simultaneously. The mean IPG lifetime was 83 +/- 14 [40-113] months. The IPG lifetime correlated with the total electrical energy delivered (P = 0.002, r = -0.496). Unilateral IPG end-of-life generally led to subacute worsening of contralateral parkinsonism. In 25% of patients, there was also a worsening of axial symptoms leading to potential medical emergencies such as falls (10%), aspiration pneumonia (10%), or psychosis (5%). A close monitoring of patients and an anticipation of IPG replacement in the case of a low-battery signal are recommended.

摘要

双侧丘脑底核(STN)刺激治疗帕金森病(PD)的疗效已得到充分证实,但对于植入式脉冲发生器(IPG)的使用寿命却知之甚少。为了研究双侧植入的Itrel II®(美敦力公司,明尼阿波利斯)脉冲发生器的使用寿命,我们回顾了在我们中心接受双侧STN慢性刺激手术的前49例连续PD患者,并记录了IPG更换前使用的刺激参数。平均电压为3.2±0.3V,平均脉冲宽度为65±10微秒,平均频率为145±16Hz。预计25%的患者会因单侧低电量信号或使用寿命结束而更换IPG。在任何一种情况下,对侧IPG都会同时更换。IPG的平均使用寿命为83±14[40-113]个月。IPG使用寿命与输送的总电能相关(P=0.002,r=-0.496)。单侧IPG使用寿命结束通常会导致对侧帕金森症状亚急性加重。在25%的患者中,还会出现轴性症状加重,导致潜在的医疗紧急情况,如跌倒(10%)、吸入性肺炎(10%)或精神病(5%)。建议密切监测患者,并在出现低电量信号时提前更换IPG。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验