使用组织模拟试验台评估动脉壁运动的组织多普勒成像测量值。

Assessment of tissue Doppler imaging measurements of arterial wall motion using a tissue mimicking test rig.

作者信息

Thrush Abigail J, Brewin Mark P, Birch Malcolm J

机构信息

Clinical Physics Department, Barts and the London NHS Trust, London.

出版信息

Ultrasound Med Biol. 2008 Mar;34(3):446-53. doi: 10.1016/j.ultrasmedbio.2007.08.011. Epub 2007 Oct 26.

Abstract

The aim of this in vitro study is to assess the accuracy of the tissue Doppler imaging arterial wall motion (TDI AWM) technique in measuring dilation over a range of distances and velocities. A test rig, consisting of two parallel blocks of tissue mimicking material (TMM), has been developed to generate known wall motion. One block remains stationary while the other moves in a cyclical motion. A calibrated laser range finder was used to measure the TMM motion. The TDI AWM measurements were found to underestimate the dilation by 21% +/- 4.7% when using the recommended scanner parameters. The size of the error was found to increase with a decrease in ultrasound output power. Results suggested that errors in the TDI AWM dilation measurements relate to underestimates in the velocity measured by the TDI technique. The error demonstrated in this study indicates a limitation in the value of TDI AWM result obtained in vivo. (E-mail: abigail.thrush@bartsandthelondon.nhs.uk).

摘要

这项体外研究的目的是评估组织多普勒成像动脉壁运动(TDI AWM)技术在测量一系列距离和速度下的扩张情况时的准确性。已开发出一种测试装置,由两块平行的组织模拟材料(TMM)组成,用于产生已知的壁运动。一块保持静止,而另一块做周期性运动。使用校准后的激光测距仪测量TMM的运动。当使用推荐的扫描仪参数时,发现TDI AWM测量低估扩张21%±4.7%。发现误差大小随超声输出功率降低而增加。结果表明,TDI AWM扩张测量中的误差与TDI技术测量的速度低估有关。本研究中证明的误差表明了体内获得的TDI AWM结果的价值存在局限性。(电子邮件:abigail.thrush@bartsandthelondon.nhs.uk

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验