To assess exposure of MR staff to static and time-varying gradient magnetic fields by direct measurement and questionnaire and to compare these to the action values and exposure limits stated in published guidance.
Instantaneous and time-averaged exposure to static magnetic fields were measured on clinical and research MR scanners. A questionnaire was sent to 100 clinical MR centers to estimate the average static field exposure. Instantaneous measurements of the time-varying magnetic field were conducted on a 1.5 T conventional and a 0.6 T open-bore MR system.
The 24-hour time-averaged static field exposure was measured to be 1.8 +/- 1 mT and the questionnaire produced a 24-hour time-average of 2.3 mT. The maximum static field strength recorded during patient set up was 554 mT. Time-varying gradient field exposure was found to exceed the action values up to 30 cm from the bore entrance or magnet rim.
The time-varying gradient field exposure limit is the only limit to be exceeded. The time-averaged static field exposure to staff was found to be 100 times lower than the recommended guidelines.
通过直接测量和问卷调查评估磁共振成像(MR)工作人员暴露于静态和随时间变化的梯度磁场中的情况,并将这些结果与已发布指南中规定的行动值和暴露限值进行比较。
在临床和研究型MR扫描仪上测量对静态磁场的瞬时和时间平均暴露。向100个临床MR中心发送问卷,以估计平均静态场暴露情况。在1.5T传统型和0.6T开放式孔径MR系统上对随时间变化的磁场进行瞬时测量。
测得24小时时间平均静态场暴露为1.8±1mT,问卷调查得出的24小时时间平均值为2.3mT。在患者安置过程中记录到的最大静态场强为554mT。发现距孔口或磁体边缘30cm范围内,随时间变化的梯度场暴露超过了行动值。
随时间变化的梯度场暴露限值是唯一被超过的限值。发现工作人员的时间平均静态场暴露比推荐指南低100倍。