Aarsand A K, Johannessen H B, Scott C S
Laboratory of Clinical Biochemistry, Haukeland University Hospital, Bergen, Norway.
Int J Lab Hematol. 2007 Dec;29(6):454-60. doi: 10.1111/j.1751-553X.2007.00912.x.
Analysis of HLA-B27 is usually performed by flow cytometry using commercial single or two colour fluorescence reagents. The CELL-DYN Sapphire (CD-Sapphire) is a high-volume routine haematology analyser that allows cell population analysis by monoclonal antibody fluorochromes analogous to flow cytometry. In this study, in-house flow cytometry analysis (n = 96, HLA-B27, One Lambda) performed on routine patient samples was used as the comparison method for analysis of HLA-B27, One Lambda (n = 40) and HLA-B27/HLA-B7, Immunotech (n = 96) reagents on the CD-Sapphire. The One Lambda results agreed 100% with the comparison method and offered clear population discrimination. The Immunotech combination also had a high level of agreement, but interpretation was more complex because of the wider cross-reactivity of the ABC-m3 antibody with B7 and other HLA-B alleles. When analysing HLA-B27 with antibodies showing nonspecific reactivity, a cut-off staining level yielding high specificity should be chosen, as the primary diagnostic value of HLA-B27 is as a 'rule-out' test for ankylosing spondylitis. The CD-Sapphire incorporates automated sampling and lysis, and medical scientists familiar with the instrument would require little additional technical training to perform the analysis. The reduced preanalytical work and total turnaround time constitute an important step towards automation of HLA-B27 and similar simple high-volume flow cytometry analysis.
HLA - B27分析通常采用流式细胞术,使用市售的单克隆或双色荧光试剂。CELL - DYN Sapphire(CD - Sapphire)是一种大容量常规血液分析仪,它可以通过类似于流式细胞术的单克隆抗体荧光染料进行细胞群体分析。在本研究中,对常规患者样本进行的内部流式细胞术分析(n = 96,HLA - B27,One Lambda)被用作在CD - Sapphire上分析One Lambda(n = 40)和Immunotech(n = 96)试剂的HLA - B27的比较方法。One Lambda的结果与比较方法100%一致,并能清晰地区分细胞群体。Immunotech组合也有很高的一致性,但由于ABC - m3抗体与B7和其他HLA - B等位基因的交叉反应性更强,其结果解读更为复杂。当使用显示非特异性反应的抗体分析HLA - B27时,应选择产生高特异性的截止染色水平,因为HLA - B27的主要诊断价值是作为强直性脊柱炎的“排除”试验。CD - Sapphire具备自动采样和裂解功能,熟悉该仪器的医学科学家几乎不需要额外的技术培训就能进行分析。减少的分析前工作和总周转时间是迈向HLA - B27自动化及类似简单大容量流式细胞术分析的重要一步。