Abstract

Two hundred three liver biopsies were performed on 114 patients suspected of having diffuse liver disease. More than one-half of the biopsies (107 of 203) were performed in liver allografts. Biopsies were performed under ultrasound (US) guidance with use of a spring-propelled 18-gauge cutting needle. Half of the biopsies were performed in patients with abnormal coagulation studies, decreased platelet counts, or both. The left lobe was chosen for biopsy over the right lobe in the majority of cases (112 vs 91). Sufficient tissue for diagnosis was obtained in 197 cases (97%). Four complications (2%) occurred that required treatment: one vasovagal reaction and three episodes of bleeding. All bleeding complications occurred in patients with coagulopathy. There was no difference in safety or efficacy between biopsies of native and transplanted livers; similarly, there was no difference in safety or efficacy between biopsies of the left and right lobe. Pain occurred less frequently with biopsies of the left lobe. US-guided biopsy with a spring-propelled 18-gauge needle is safe and efficacious in the diagnosis of diffuse liver disease.

摘要

对114例疑似患有弥漫性肝病的患者进行了203次肝脏活检。超过一半的活检(203例中的107例)是在肝移植患者中进行的。活检在超声(US)引导下,使用弹簧驱动的18号切割针进行。一半的活检是在凝血检查异常、血小板计数减少或两者兼有的患者中进行的。在大多数情况下(112例对91例),选择左叶而非右叶进行活检。197例(97%)获得了足够用于诊断的组织。发生了4例(2%)需要治疗的并发症:1例血管迷走神经反应和3次出血事件。所有出血并发症均发生在凝血病患者中。原发性肝脏活检和移植肝脏活检在安全性或有效性方面没有差异;同样,左叶活检和右叶活检在安全性或有效性方面也没有差异。左叶活检时疼痛发生频率较低。超声引导下使用弹簧驱动的18号针进行活检在弥漫性肝病的诊断中是安全有效的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验