Computed tomography phantom for radiochromic film dosimetry.

作者信息

Gotanda R, Katsuda T, Gotanda T, Eguchi M, Takewa S, Tabuchi A, Yatake H

机构信息

Graduate School of Health Sciences, Okayama University, Okayama, Japan.

出版信息

Australas Phys Eng Sci Med. 2007 Sep;30(3):194-9. doi: 10.1007/BF03178426.

Abstract

To evaluate in detail the dose distribution during computed tomography (CT), a sheet roll CT dosimetry phantom (SRCT-P) with a radiochromic film (RF) was experimentally developed. The SRCT-P was made by rolling up a vinyl chloride sheet in a cylindrical shape to arbitrarily select the SRCT-P diameter, dose measurement position, and depth. The SRCT-P centre core consisted of a plastic hose in which a 10 mm acrylic bar with a RF was inserted. To determine the availability of the SRCT-P, the surface and centre doses (at a 5 mm radius) at each SRCT-P diameter (6-16 cm; every 2 cm) were measured. The ratios of the centre-to-surface doses (D(centre)/D(surface)) systematically increased, from 80 to 111%, for decreasing SRCT-P diameters, between 16 and 6 cm, respectively. The centre dose approached the surface dose as the SRCT-P diameter decreased. To use a RF for a CT dose measurement, further detailed research and analysis is necessary. However, this study has shown that a SRCT-P is useful and beneficial for the measurement of the dose distribution during a CT examination.