Büsing Karen A, Kilian A Kristina, Schaible Thomas, Debus Angelika, Weiss Christel, Neff K Wolfgang
Department of Clinical Radiology, University Hospital Mannheim, University of Heidelberg, Theodor Kutzer Ufer 1-3, 68167 Mannheim, Germany.
Radiology. 2008 Feb;246(2):553-61. doi: 10.1148/radiol.2462062166. Epub 2007 Nov 30.
To prospectively assess the reliability of magnetic resonance (MR) image volume measurement in fetuses with congenital diaphragmatic hernia (CDH) and the reliability and validity of measurements in in vitro lung models.
This study was approved by the ethics committee, and informed consent was obtained. MR fetal lung volume (FLV) was measured in 40 consecutive fetuses with CDH by using half-Fourier acquired single-shot turbo spin-echo MR imaging and true fast imaging with steady-state precession at 24-36 weeks gestation (mean gestational age, 30.6 weeks +/- 3.5 [standard deviation]). Lung volumes were independently assessed in three orthogonal section planes by two experienced observers. Additionally, 28 in vitro lung models of defined volumes of 1-60 mL were evaluated the same way. To assess measurement validity and reliability, the intraclass correlation coefficient (ICC) and the Bland-Altman plot were used.
The interobserver reliability was high for both the lung models and FLV measurements (ICC, 0.999 and 0.928, respectively). Measurement validity was also good, with a mean difference between the calculated volume and the true volume of 0.4 mL (95% confidence interval: 0.30, 0.48). Measurement reliability and validity did not depend, to any considerable degree, on imaging plane or sequence (ICC range, 0.878-0.999) or on total volume.
The reliability and validity of MR volume measurements are high. The method is independent of the sequence and the imaging plane and can be performed with a very good interobserver agreement, even in small volumes.
前瞻性评估先天性膈疝(CDH)胎儿磁共振(MR)图像体积测量的可靠性以及体外肺模型测量的可靠性和有效性。
本研究经伦理委员会批准,并获得知情同意。在40例连续的CDH胎儿中,于妊娠24 - 36周(平均孕周30.6周±3.5[标准差])采用半傅里叶采集单次激发快速自旋回波MR成像和稳态进动快速成像测量胎儿肺体积(FLV)。由两名经验丰富的观察者在三个相互垂直的截面平面上独立评估肺体积。此外,对28个定义体积为1 - 60 mL的体外肺模型采用相同方法进行评估。为评估测量的有效性和可靠性,使用组内相关系数(ICC)和Bland - Altman图。
肺模型和FLV测量的观察者间可靠性均较高(ICC分别为0.999和0.928)。测量有效性也良好,计算体积与真实体积的平均差异为0.4 mL(95%置信区间:0.30,0.48)。测量可靠性和有效性在很大程度上不依赖于成像平面或序列(ICC范围为0.878 - 0.999)或总体积。
MR体积测量的可靠性和有效性较高。该方法独立于序列和成像平面,即使在小体积情况下也能获得很好的观察者间一致性。