用于测量生殖器压力疼痛阈值的外阴压力计。

The vulvalgesiometer as a device to measure genital pressure-pain threshold.

作者信息

Pukall Caroline F, Young Roy A, Roberts Michael J, Sutton Katherine S, Smith Kelly B

机构信息

Department of Psychology, Queen's University, Humphrey Hall, 62 Arch Street, Kingston, Ontario, K7L 3N6, Canada.

出版信息

Physiol Meas. 2007 Dec;28(12):1543-50. doi: 10.1088/0967-3334/28/12/008. Epub 2007 Dec 3.

Abstract

The construction and application of the vulvalgesiometer are described. This manually-applied device allows for the quantifiable measurement of pressure-pain thresholds in the external female genital region. A set of five vulvalgesiometers exerting pressures from 3 to 950 g was used in two studies. The goal of the first study was to examine the ability of the vulvalgesiometers to discriminate between women with and without provoked vestibulodynia (PVD). In a matched sample of affected and non-affected women, women with PVD exhibited significantly lower vestibular pressure-pain thresholds as compared to control women. As well, approximately half the sample of women with PVD described the sensation elicited at pressure-pain threshold as similar to the pain experienced during sexual intercourse. The goal of the second study was to investigate the inter-rater reliability of the vulvalgesiometers. In this separate sample of women with and without PVD, each participant was tested for pressure-pain threshold by two different investigators at different times. Results demonstrated high levels of inter-rater reliability, indicating that the vulvalgesiometers can be consistently used by different investigators. Further, results indicated significant negative correlations between pressure-pain thresholds and pain intensity ratings recorded during the cotton-swab test, suggesting that the lower the threshold, the higher the pain ratings during vestibular palpation. The vulvalgesiometers can be utilized for several purposes, including treatment outcome studies and measuring the degree of PVD severity.

摘要

本文描述了外阴压力痛觉计的构建与应用。这种手动操作的设备能够对外阴女性生殖区域的压力疼痛阈值进行可量化测量。在两项研究中使用了一组五个能施加3至950克压力的外阴压力痛觉计。第一项研究的目的是检验外阴压力痛觉计区分患有和未患有激发性前庭疼痛障碍(PVD)的女性的能力。在患病人群与非患病人群相匹配的样本中,与对照女性相比,患有PVD的女性前庭压力疼痛阈值显著更低。此外,约一半患有PVD的女性样本描述在压力疼痛阈值时引发的感觉类似于性交时经历的疼痛。第二项研究的目的是调查外阴压力痛觉计的评分者间信度。在这个患有和未患有PVD的女性的独立样本中,每位参与者在不同时间由两名不同的研究者测试压力疼痛阈值。结果显示评分者间信度水平很高,表明不同研究者能够一致地使用外阴压力痛觉计。此外,结果表明压力疼痛阈值与棉签试验期间记录的疼痛强度评分之间存在显著负相关,这表明阈值越低,前庭触诊时的疼痛评分越高。外阴压力痛觉计可用于多种目的,包括治疗效果研究以及测量PVD的严重程度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验