Koziak Adrian M, Winter Jeff, Lee Ting-Yim, Thompson R Terry, St Lawrence Keith S
Lawson Health Research Institute, University of Western Ontario, London, Ontario, Canada N6A 4V2.
Magn Reson Imaging. 2008 May;26(4):543-53. doi: 10.1016/j.mri.2007.10.005. Epub 2007 Dec 11.
Abnormalities in cerebral blood flow (CBF) are believed to play a significant role in the development of major neonatal neuropathologies. One approach that would appear ideal for measuring CBF in this fragile age group is arterial spin labeling (ASL) since ASL techniques are noninvasive and quantitative. The purpose of this study was to assess the accuracy of a pulsed ASL method implemented on a 3-T scanner dedicated to neonatal imaging. Cerebral blood flow was measured in nine neonatal piglets, the ASL-CBF measurements were acquired at two inversion times (TI) (1,200 and 1,700 ms), and CBF was measured by perfusion computed tomography (pCT) for validation. Perfusion CT also provided images of cerebral blood volume, which were used to identify large blood vessels, and contrast arrival time, which were used to assess differences in arterial transit times between gray and white matter. Good agreement was found between gray matter CBF values from pCT (76+/-1 ml/min per 100 g) and ASL at TI=1,700 ms (73+/-1 ml/min per 100 g). At TI=1,200 ms, ASL overestimated CBF (91+/-2 ml/min per 100 g), which was attributed to substantial intravascular signal. No significant differences in white matter CBF from pCT and ASL were observed (average CBF=60+/-1 ml/min per 100 g), nor was there any difference in contrast arrival times for gray and white matter (0.95+/-0.04 and 0.99+/-0.03 s, respectively), which suggests that the arterial transit times for ASL were the same in this animal model. This study verified the accuracy of the implemented ASL technique and showed the value of using pCT to study other factors that can affect ASL-CBF measurements.
脑血流量(CBF)异常被认为在主要新生儿神经病理学的发展中起重要作用。对于在这个脆弱的年龄组中测量CBF而言,一种看似理想的方法是动脉自旋标记(ASL),因为ASL技术是非侵入性的且可定量。本研究的目的是评估在一台专门用于新生儿成像的3-T扫描仪上实施的脉冲ASL方法的准确性。对9只新生仔猪进行了脑血流量测量,在两个反转时间(TI)(1200和1700毫秒)获取ASL-CBF测量值,并通过灌注计算机断层扫描(pCT)测量CBF以进行验证。灌注CT还提供了脑血容量图像,用于识别大血管,以及对比剂到达时间,用于评估灰质和白质之间动脉传输时间的差异。发现pCT灰质CBF值(每100克76±1毫升/分钟)与TI = 1700毫秒时的ASL值(每100克73±1毫升/分钟)之间具有良好的一致性。在TI = 1200毫秒时,ASL高估了CBF(每100克91±2毫升/分钟),这归因于大量血管内信号。未观察到pCT和ASL在白质CBF上有显著差异(平均CBF =每100克60±1毫升/分钟),灰质和白质的对比剂到达时间也没有差异(分别为0.95±0.04和0.99±0.03秒),这表明在该动物模型中ASL的动脉传输时间相同。本研究验证了所实施的ASL技术的准确性,并显示了使用pCT研究其他可能影响ASL-CBF测量的因素的价值。