心脏多层螺旋CT造影剂注射方案的优化:主团注造影剂密度对测试团注参数及患者体重的依赖性

Optimization of cardiac MSCT contrast injection protocols: dependency of the main bolus contrast density on test bolus parameters and patients' body weight.

作者信息

Rist Carsten, Becker Christoph R, Kirchin Miles A, Johnson Thorsten R, Busch Stefanie, Bae Kyongtae Ty, Leber Alexander W, Reiser Maximilian F, Nikolaou Konstantin

机构信息

Department of Clinical Radiology, University Hospitals - Grosshadern, Ludwig-Maximilians University, Marchioninistr. 15, 81377 Munich, Germany.

出版信息

Acad Radiol. 2008 Jan;15(1):49-57. doi: 10.1016/j.acra.2007.08.005.

Abstract

RATIONALE AND OBJECTIVES

Our aim was to evaluate the correlation of test bolus (TB) curve parameters with main bolus (MB) contrast density for cardiac 16-slice computed tomography, and to correlate observed enhancement with patient body weight.

MATERIALS AND METHODS

Sixty patients with known or suspected coronary artery disease were included in a prospective double-blind study. Contrast material containing 300 mg iodine/mL (Iomeprol 300; Imeron 300, Bracco Imaging SpA, Milan, Italy) and 400 mg iodine/mL (Iomeprol 400; Imeron 400) was injected at a rate of 1 g of iodine/second. Contrast densities (Hounsfield units) of the MB were determined in the left cardiac system. The peak density (PD) of maximum attenuation and the area under the curve (AUC) of the TB curve were calculated for each patient. The dependency of MB contrast attenuation on these parameters and on patient body weight was evaluated.

RESULTS

Positive correlations (r = 0.52 and r = 0.56, respectively; P < .0001) were obtained between the PD and AUC of the TB curve with the mean density of the MB. Stronger correlations (r = 0.63 and r = 0.64, respectively; P < .0001) between PD and AUC of the TB curve and MB attenuation were found when patient body weight was included in the analysis.

CONCLUSIONS

Strong correlation of the PD and AUC of the TB curve with the mean density of the MB is observed when patient body weight is considered. Contrast injection protocols may be optimized, and variations of MB contrast density in the left ventricle and main coronary arteries reduced, by taking these TB parameters and the weight of the patient into account.

摘要

原理与目的

我们的目的是评估心脏16层计算机断层扫描中测试团注(TB)曲线参数与主团注(MB)对比剂密度之间的相关性,并将观察到的强化与患者体重相关联。

材料与方法

60例已知或疑似冠心病患者纳入一项前瞻性双盲研究。以1克碘/秒的速率注射含300毫克碘/毫升(碘海醇300;欧乃影300,意大利米兰博莱科影像股份公司)和400毫克碘/毫升(碘海醇400;欧乃影400)的对比剂。在左心系统中测定MB的对比剂密度(亨氏单位)。计算每位患者TB曲线的最大衰减峰值密度(PD)和曲线下面积(AUC)。评估MB对比剂衰减对这些参数以及患者体重的依赖性。

结果

TB曲线的PD和AUC与MB的平均密度之间呈正相关(分别为r = 0.52和r = 0.56;P <.0001)。当分析中纳入患者体重时,TB曲线的PD和AUC与MB衰减之间的相关性更强(分别为r = 0.63和r = 0.64;P <.0001)。

结论

当考虑患者体重时,观察到TB曲线的PD和AUC与MB的平均密度之间存在强相关性。通过考虑这些TB参数和患者体重,可以优化对比剂注射方案,并减少左心室和主要冠状动脉中MB对比剂密度的变化。