CE coupled with amperometric detection using a boron-doped diamond microelectrode: validation of a method for endogenous norepinephrine analysis in tissue.

作者信息

Quaiserová-Mocko Veronika, Novotný Martin, Schaefer Luther S, Fink Gregory D, Swain Greg M

机构信息

Department of Chemistry, Michigan State University, East Lansing, MI, USA.

出版信息

Electrophoresis. 2008 Jan;29(2):441-7. doi: 10.1002/elps.200700398.

Abstract

The level of endogenous norepinephrine (NE) in several tissue types was determined by CE with amperometric detection. We report herein on the method validation by HPLC using both amperometric and coulometric detection (CD). Keys to the method were the use of a diamond microelectrode for detection and off-line SPE for sample preparation. The run buffer was a 250 mM borate solution adjusted to pH 8.8 with potassium hydroxide. The diamond microelectrode exhibited a low and stable background current, and a low peak-to-peak noise < or =0.65 pA at the detection potential of +0.86 V versus Ag/AgCl. For standard solutions, the detector signal (i.e., oxidation current) changed linearly with the NE concentration (r(2) = 0.999) between 60 and 1000 nmol/L with an estimated LOD of 51 nmol/L (S/N = 3) and a response variability of 4.5% (RSD, n = 5). An Oasis MCX sorbent was used for SPE and the procedure produced an NE recovery of 95.1 +/- 5.6% (n = 6) from tissue homogenates. NE levels in the spleen, small intestine, and heart of a normotensive rat were found to be in the range of 0.77-0.97, 0.22-0.32, and 0.29-0.45 microg/g tissue (n = 3), respectively.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验