Enhancers of KCNQ channels are known to be effective in chronic pain models. To discover novel enhancers of KCNQ channels, the authors developed a medium-throughput electrophysiological assay by using the IonWorks platform. Screening of 20 CHO-K1 clones stably expressing KCNQ2/3 was performed on the IonWorks HT until the best clone (judged from seal rate, current level, and stability) was obtained. The KCNQ2/3 current amplitude in the cells was found to increase from 60 +/- 15 pA to 473 +/- 80 pA (at -10 mV), and the expression rate was increased by 56% when the cells were incubated at 27 degrees C overnight. The clone used for compound screening had a seal rate of greater than 90% and an overall success rate of greater than 70%. The voltage step protocol (hold cells at -80 mV and depolarize to -10 mV for 1 s) was designed to provide moderate current but still allow for pharmacological current enhancement. EC(50)s were generated from 8-point concentration-response curves with a control compound on each plate using compounds that were also tested with conventional patch clamp. The authors found that there was a very good correlation (R(2) > 0.9) between the 2 assays, thus demonstrating the highly predictive nature of the IonWorks assay.

摘要

已知钾离子通道(KCNQ)增强剂在慢性疼痛模型中有效。为了发现新型KCNQ通道增强剂,作者利用IonWorks平台开发了一种中等通量的电生理检测方法。在IonWorks HT上对20个稳定表达KCNQ2/3的中国仓鼠卵巢细胞(CHO-K1)克隆进行筛选,直到获得最佳克隆(根据封接率、电流水平和稳定性判断)。当细胞在27摄氏度下孵育过夜时,细胞中的KCNQ2/3电流幅度从60±15 pA增加到473±80 pA(在-10 mV时),表达率提高了56%。用于化合物筛选的克隆封接率大于90%,总体成功率大于70%。电压阶跃方案(将细胞保持在-80 mV并去极化至-10 mV持续1秒)旨在提供适度电流,但仍允许药理学电流增强。使用也用传统膜片钳检测的化合物,从每个平板上与对照化合物的8点浓度-反应曲线生成半数有效浓度(EC50)。作者发现这两种检测方法之间有非常好的相关性(R2>0.9),从而证明了IonWorks检测方法具有高度预测性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验