脐动脉多普勒波形质量指标的临床试验

Clinical trial of umbilical artery Doppler waveform quality indices.

作者信息

Hoskins P R

机构信息

Medical Physics and Medical Engineering Department, Royal Infirmary, Edinburgh, UK.

出版信息

Ultrasound Med Biol. 1991;17(8):777-86. doi: 10.1016/0301-5629(91)90160-x.

Abstract

The ability of two Doppler waveform quality indices to discriminate between high- and low-quality waveforms was tested using 427 sets of umbilical artery Doppler waveforms from patients. The waveforms had been acquired using a 4-MHz continuous-wave Doppler unit. The quality indices (QI) were based on an assessment of the degree of noise of the maximum frequency envelope of the waveforms, and were first a correlation between successive waveform envelopes (QI1), and, second, a sum of local linearity measures (QI2). The sets of waveforms were graded subjectively according to the clarity of the outline of the waveforms, the degree of interference in the region of the spectrum above the outline, and in terms of the degree of variability caused by fetal breathing. At 90% sensitivity for detection of low-quality waveforms according to a high envelope clarity score, the specificities were 68.2% and 52.7%, respectively, for QI1 and QI2. QI1 was independent from pulsatility index and waveform length, but showed strong dependence on fetal breathing. QI2 showed strong independence from pulsatility and fetal breathing and reasonable independence from waveform length. Both QI1 and QI2 performed poorly when there was a large degree of noise in the region of the spectrum above the envelope; however, this poor performance was often related to the inability of the maximum frequency follower to estimate correctly the maximum frequency envelope in those conditions so that the high waveform quality values reflected the erroneous calculation of pulsatility index in those cases.

摘要

使用来自患者的427组脐动脉多普勒波形,测试了两种多普勒波形质量指标区分高质量和低质量波形的能力。这些波形是使用4兆赫连续波多普勒装置采集的。质量指标(QI)基于对波形最大频率包络噪声程度的评估,首先是连续波形包络之间的相关性(QI1),其次是局部线性测量值的总和(QI2)。根据波形轮廓的清晰度、轮廓上方频谱区域的干扰程度以及胎儿呼吸引起的变异程度,对波形集进行主观分级。根据高包络清晰度评分,在检测低质量波形的灵敏度为90%时,QI1和QI2的特异性分别为68.2%和52.7%。QI1与搏动指数和波形长度无关,但对胎儿呼吸有很强的依赖性。QI2与搏动和胎儿呼吸有很强的独立性,与波形长度有合理的独立性。当包络上方频谱区域存在大量噪声时,QI1和QI2的表现都很差;然而,这种不佳表现通常与最大频率跟踪器在这些情况下无法正确估计最大频率包络有关,因此在这些情况下,高波形质量值反映了搏动指数的错误计算。

相似文献

6
The choice of index for umbilical artery Doppler waveforms.脐动脉多普勒波形指标的选择。
Ultrasound Med Biol. 1989;15(2):107-11. doi: 10.1016/0301-5629(89)90159-2.
7
Determination of the presence of fetal apnoea using umbilical artery and umbilical vein Doppler waveforms.
Ultrasound Med Biol. 1988;14(7):589-92. doi: 10.1016/0301-5629(88)90126-3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验