Ramcharitar Steve, Daeman Joost, Patterson Mark, van Guens Robert Jan, Boersma Eric, Serruys Patrick W, van der Giessen Willem J
Thoraxcenter, Erasmus Medical Center, Rotterdam, The Netherlands.
Catheter Cardiovasc Interv. 2008 Jan 1;71(1):44-50. doi: 10.1002/ccd.21418.
The in vivo comparison of the accuracy of two 3-dimensional quantitative coronary angiography (QCA) systems.
Precision-drilled plexiglass phantoms with five different luminal diameters (0.5-1.9 mm) were percutaneously inserted into the coronary arteries of four Yorkshire pigs. Twenty-one angiographic images of these stenotic phantoms were acquired for in vivo validation testing. Quantitative assessments of the minimum, maximum, and mean luminal diameters together with the minimum luminal area were determined using two 3D QCA systems, the CardiOp-B and CAAS 5.
The CardiOp-B system significantly underestimated the minimum luminal diameter MLD whilst both systems significantly overestimated the maximum luminal diameter at the minimal luminal area (MLA) over the phantom's true value. The CAAS 5 system had a greater degree of accuracy/mm (mean difference = 0.01 vs. 0.03) and precision/mm (SD = 0.09 vs. 0.23) than the CardiOp-B in assessing the minimal LD. An increased precision/mm (SD = 0.01 vs. 0.29) and accuracy/mm (mean difference = 0.03 vs. 0.11) in the mean LD was observed with the CAAS 5. In comparing the MLA/mm(2) the CAAS 5 was more precise/mm(2) (SD = 0.14 vs. 0.55) and accurate/mm(2) (mean difference = 0.12 vs. 0.02) to the true phantom MLA compared to the CardiOp-B system.
In a 21 phantom study, the CAAS 5 3D QCA system had a greater degree of accuracy and precision in both the luminal and area measurements than the CardiOp-B 3D QCA system.
对两种三维定量冠状动脉造影(QCA)系统的准确性进行体内比较。
将具有五种不同管腔直径(0.5 - 1.9毫米)的精密钻孔有机玻璃模型经皮插入四只约克夏猪的冠状动脉。获取这些狭窄模型的21张血管造影图像用于体内验证测试。使用两种三维QCA系统,即CardiOp - B和CAAS 5,对最小、最大和平均管腔直径以及最小管腔面积进行定量评估。
CardiOp - B系统显著低估了最小管腔直径(MLD),而在最小管腔面积(MLA)处,两种系统均显著高估了最大管腔直径,超过了模型的真实值。在评估最小管腔直径时,CAAS 5系统比CardiOp - B系统具有更高的每毫米准确度(平均差异 = 0.01对0.03)和每毫米精密度(标准差 = 0.09对0.23)。使用CAAS 5时,观察到平均管腔直径的每毫米精密度(标准差 = 0.01对0.29)和每毫米准确度(平均差异 = 0.03对0.11)有所提高。在比较每平方毫米的MLA时,与CardiOp - B系统相比,CAAS 5在每平方毫米上更精确(标准差 = 0.14对0.55)且更准确(平均差异 = 0.12对0.02),与真实模型的MLA相比。
在一项包含21个模型的研究中,CAAS 5三维QCA系统在管腔和面积测量方面比CardiOp - B三维QCA系统具有更高的准确度和精密度。