Suppr超能文献

动脉瘤夹闭术后非螺旋采集CT血管造影:体外测试显示伪影减少。

Nonhelical acquisition CT angiogram after aneurysmal clipping: in vitro testing shows diminished artifact.

作者信息

Mamourian A C, Erkmen K, Pluta D J

机构信息

Department of Radiology, Dartmouth-Hitchcock Medical Center, Lebanon, NH 03756, USA.

出版信息

AJNR Am J Neuroradiol. 2008 Apr;29(4):660-2. doi: 10.3174/ajnr.A0893. Epub 2008 Jan 9.

Abstract

Imaging of the cerebral vessels with use of CT angiogram (CTA) after placement of aneurysmal clips is often limited by clip artifacts. We used a phantom to demonstrate a visible reduction in metal artifact when using the axial technique, compared with the usual CTA helical acquisition. This approach may have some advantage when used for CTA with 64-section scanners in the specific circumstance of immediate postoperative imaging after placement of cerebral aneurysmal clips.

摘要

动脉瘤夹置入后使用CT血管造影(CTA)对脑血管进行成像时,往往会受到夹子伪影的限制。我们使用了一个模型来证明,与常规CTA螺旋采集相比,采用轴向技术时金属伪影明显减少。在使用64层扫描仪进行CTA检查时,这种方法在脑动脉瘤夹置入术后立即成像的特定情况下可能具有一些优势。

相似文献

1
Nonhelical acquisition CT angiogram after aneurysmal clipping: in vitro testing shows diminished artifact.
AJNR Am J Neuroradiol. 2008 Apr;29(4):660-2. doi: 10.3174/ajnr.A0893. Epub 2008 Jan 9.
2
Computed tomographic angiography in the evaluation of clip placement for intracranial aneurysm.
Acta Neurochir (Wien). 2008 Jul;150(7):669-76. doi: 10.1007/s00701-008-1515-z. Epub 2008 May 21.
5
Vascular assessment after clipping surgery using four-dimensional CT angiography.
Neurosurg Rev. 2019 Mar;42(1):107-114. doi: 10.1007/s10143-018-0962-0. Epub 2018 Mar 3.

引用本文的文献

3
Quantitative analysis of factors affecting cobalt alloy clip artifacts in computed tomography.
J Korean Neurosurg Soc. 2014 Nov;56(5):400-4. doi: 10.3340/jkns.2014.56.5.400. Epub 2014 Nov 30.
4
The value of dual-energy CTA for control of surgically clipped aneurysms.
Eur Radiol. 2011 Oct;21(10):2193-201. doi: 10.1007/s00330-011-2147-x. Epub 2011 May 10.
5
[Modern cross-sectional imaging of head and neck vessels].
Radiologe. 2010 Apr;50(4):377-98. doi: 10.1007/s00117-009-1974-0.

本文引用的文献

3
Artifacts in CT: recognition and avoidance.
Radiographics. 2004 Nov-Dec;24(6):1679-91. doi: 10.1148/rg.246045065.
4
Intraoperative angiography during aneurysm surgery: a prospective evaluation of efficacy.
J Neurosurg. 2002 Jun;96(6):993-9. doi: 10.3171/jns.2002.96.6.0993.
5
Late angiographic follow-up review of surgically treated aneurysms.
J Neurosurg. 1999 Sep;91(3):396-401. doi: 10.3171/jns.1999.91.3.0396.
6
Titanium aneurysm clips: Part I--Mechanical, radiological, and biocompatibility testing.
Neurosurgery. 1996 Jun;38(6):1158-63; discussion 1164. doi: 10.1097/00006123-199606000-00022.