钛动脉瘤夹:第一部分——力学、放射学及生物相容性测试。

Titanium aneurysm clips: Part I--Mechanical, radiological, and biocompatibility testing.

作者信息

Lawton M T, Ho J C, Bichard W D, Coons S W, Zabramski J M, Spetzler R F

机构信息

Division of Neurological Surgery, St. Joseph's Hospital and Medical Center, Phoenix, Arizona, USA.

出版信息

Neurosurgery. 1996 Jun;38(6):1158-63; discussion 1164. doi: 10.1097/00006123-199606000-00022.

Abstract

Most aneurysm clips are made of cobalt-based alloys. Although these clips are nonferromagnetic, they still produce artifact that degrades the quality of magnetic resonance (MR) images. A new aneurysm clip of pure titanium was developed to minimize artifact on postoperative MR images. We evaluated these clips in a series of mechanical tests in vitro, biocompatibility tests in rabbits, and radiological tests in greyhound dogs. The clip sizes and shapes matched those of conventional aneurysm clips. The average closing forces ranged between 151.6 and 181.8 g and were not diminished by repeated sterilization or stress. After > 20 million cycles of high-pressure and high-frequency pulsations, the clips did not open and the closing forces were not reduced. Titanium aneurysm clips implanted in the subarachnoid space of 12 rabbits for 1 or 6 months produced mild gliosis identical to that produced by implantation of cobalt alloy clips in 12 control rabbits. Based on pre- and postoperative weights and electron microscopic scans, the titanium implants did not corrode. The artifact on computed tomographic and MR imaging produced by a titanium clip placed on the internal carotid artery of a greyhound was less than that produced by an identical cobalt-chrome alloy clip by a factor of two to three. This study demonstrated that titanium aneurysm clips are mechanically equivalent to conventional clips, biocompatible, and corrosion resistant. Furthermore, titanium clips have superior imaging characteristics, creating less computed tomographic and MR imaging artifact and permitting better resolution of anatomic structures than cobalt alloy clips.

摘要

大多数动脉瘤夹由钴基合金制成。尽管这些夹子是非铁磁性的,但它们仍会产生伪影,降低磁共振(MR)图像的质量。一种新型的纯钛动脉瘤夹被研发出来,以尽量减少术后MR图像上的伪影。我们在一系列体外力学测试、家兔生物相容性测试以及灵缇犬放射学测试中对这些夹子进行了评估。夹子的尺寸和形状与传统动脉瘤夹相匹配。平均闭合力量在151.6至181.8克之间,且不会因反复消毒或施加应力而减弱。在经历超过2000万次的高压和高频脉动循环后,夹子未张开,闭合力量也未降低。将钛动脉瘤夹植入12只家兔的蛛网膜下腔1个月或6个月,所产生的轻度胶质增生与将钴合金夹植入12只对照家兔所产生的相同。根据术前和术后的体重以及电子显微镜扫描结果,钛植入物未发生腐蚀。放置在灵缇犬颈内动脉上的钛夹在计算机断层扫描和MR成像上产生的伪影比相同的钴铬合金夹产生的伪影少两到三倍。这项研究表明,钛动脉瘤夹在力学性能上与传统夹子相当,具有生物相容性且耐腐蚀。此外,钛夹具有优越的成像特性,产生的计算机断层扫描和MR成像伪影更少,与钴合金夹相比能更好地分辨解剖结构。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验