Three-dimensional finite element stress analysis in and around the Screw-Vent implant.

作者信息

Clelland N L, Ismail Y H, Zaki H S, Pipko D

机构信息

Ohio State University College of Dentistry, Columbus.

出版信息

Int J Oral Maxillofac Implants. 1991 Winter;6(4):391-8.

PMID:1820307
Abstract

The three-dimensional finite element stress analysis method was used to determine the pattern and concentration of stresses within the Screw-Vent endosseous implant and its supporting tissues. For this commercially pure titanium implant, maximum stresses were located within the implant collar immediately below the bony crest. These stresses were at least 18 times less than the endurance limit of commercially pure titanium (259.90 MPa). Maximum stresses (19.57 MPa) in the bone were lingual to the superior portion of the collar. Previous longitudinal radiographic studies of a similar implant have revealed bone loss mesial and distal to the implant. For the Screw-Vent, mesial and distal stresses (maximum 0.38 MPa) were much lower than those buccal and lingual to the implant. The clinical significance of the stress transfer to the bone buccal and lingual to the implant has yet to be determined.