Montreal Neurological Inst., McGill Univ., Que.
IEEE Trans Med Imaging. 1991;10(1):90-8. doi: 10.1109/42.75615.
Initial experience is reported with the Scanditronix PC 2048-15B, a 15-slice positron emission tomography (PET) system using multicrystal/multiphoto-multiplier modules to obtain high spatial resolution. Random and scattered events are reduced using an orbiting (68)Ge rod source for transmission scans by only accepting coincidence lines which intersect the instantaneous position of the source. Scatter correction of the emission data is removed with a deconvolution kernel, random and dead-time correction by the use of observed singles rates. The peak count rates are 11.7/20.0 Kcps for the direct cross slices at concentrations of 4.5/5.1 muCi/cc. respectively. Over the radial range 0-9 cm from the ring center, radial transverse resolution is 4.6-6.4 mm, aid tangential transverse resolution is 4.6-5.1 mm using a Hanning filter. Over the same range, axial resolution varies from 6.1-6.2 mm in direct slices and from 5.4-7.1 mm in cross slices. This near-isotropic resolution allows collection of image volume data with no preferred direction for signal averaging errors.
本文报道了 Scanditronix PC 2048-15B 的初步经验,这是一种 15 层正电子发射断层扫描(PET)系统,使用多晶体/多光电倍增管模块获得高空间分辨率。随机和散射事件通过仅接受与源的瞬时位置相交的符合线,使用轨道(68)Ge 棒源进行传输扫描来减少。通过使用观察到的单粒子率,使用反卷积核去除发射数据的散射校正,随机和死时间校正。在浓度分别为 4.5/5.1 μCi/cc 时,直接交叉切片的峰值计数率分别为 11.7/20.0 Kcps。在距环中心 0-9 cm 的径向范围内,径向横向分辨率为 4.6-6.4 mm,使用汉宁滤波器时,切向横向分辨率为 4.6-5.1 mm。在相同的范围内,直接切片的轴向分辨率从 6.1-6.2mm 变化,而交叉切片的轴向分辨率从 5.4-7.1mm 变化。这种近各向同性分辨率允许在没有信号平均误差首选方向的情况下采集图像体积数据。