Aguinaga N, Campillo N, Viñas P, Hernández-Córdoba M
Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, 30071, Murcia, Spain.
Anal Bioanal Chem. 2008 Jun;391(4):1419-24. doi: 10.1007/s00216-008-1862-x. Epub 2008 Feb 3.
This paper describes a headspace solid-phase microextraction (HS-SPME) procedure coupled to gas chromatography with mass spectrometric detection (GC-MS) for the determination of eight PAHs in aquatic species. The influence of various parameters on the PAH extraction efficiency was carefully examined. At 75 degrees C and for an extraction time of 60 min, a polydimethylsiloxane-divinylbenzene (PDMS/DVB) fiber coating was found to be most suitable. Under the optimized conditions, detection limits ranged from 8 to 450 pg g(-1), depending on the compound and the sample matrix. The repeatability varied between 7 and 15% (RSD). Accuracy was tested using the NIST SRM 1974b reference material. The method was successfully applied to different samples, and the studied PAHs were detected in several of the samples.
本文描述了一种顶空固相微萃取(HS-SPME)与气相色谱-质谱检测(GC-MS)联用的方法,用于测定水生物种中的8种多环芳烃。仔细研究了各种参数对多环芳烃萃取效率的影响。在75℃和60分钟的萃取时间下,发现聚二甲基硅氧烷-二乙烯基苯(PDMS/DVB)纤维涂层最为合适。在优化条件下,检测限为8至450 pg g(-1),具体取决于化合物和样品基质。重复性在7%至15%(相对标准偏差)之间变化。使用NIST SRM 1974b标准物质测试了准确性。该方法成功应用于不同样品,并且在多个样品中检测到了所研究的多环芳烃。