Cueva-Mestanza R, Sosa-Ferrera Z, Torres-Padrón M E, Santana-Rodríguez J J
Department of Chemistry, University of Las Palmas de Gran Canaria, 35017 Las Palmas de Gran Canaria, Spain.
J Chromatogr B Analyt Technol Biomed Life Sci. 2008 Feb 15;863(1):150-7. doi: 10.1016/j.jchromb.2008.01.026. Epub 2008 Jan 20.
An analytical method combining microwave assisted micellar extraction (MAME) and solid phase extraction (SPE) has been developed to extract and preconcentrate a selected group of eight pharmaceutical compounds in sediment samples prior to their determination using liquid chromatography with an UV-DAD detector. A non-ionic surfactant, Polyoxyethylene 10 lauryl ether (POLE) was used for the MAME extraction and the different parameters for the optimization process were studied. Then, SPE was used to clean-up and preconcentrate the target analytes in the extract, prior to their determination using HPLC-UV. The method was applied to the determination of the selected pharmaceuticals compounds in several kinds of sediment samples with different characteristics. Relative recoveries for spiked sediment samples were over 70% and relative standard deviations (RSDs) were under 11% for all recoveries tested. Detection limits between 4 and 167ngg(-1) were obtained. The method was validated using Soxhlet extraction procedure.