A reversed-phase isocratic high-performance liquid chromatographic method for the determination of total teicoplanin in plasma is reported. The method developed uses a bracketing injection technique in conjunction with large injection volumes on a 1 mm diameter column to form a limited injection-generated gradient. The chromatography yields adequate resolution among all the major components for individual quantitation and also allows quantitation of total teicoplanin in plasma using ultraviolet detection. Pretreatment is by solid-phase extraction which uses C8 Bond Elut cartridges and gives effective clean up from endogenous materials. The method offers a faster and simplified means to determine total teicoplanin in plasma than those previously reported, and has a detection limit of 50 ng/ml.