Suppr超能文献

基于丙酮和乙醇的双水相萃取后荧光法测定痕量维生素B2

[Determination of trace vitamin B2 by fluorimetry after aqueous two-phase extraction based on acetone and ethanol].

作者信息

Long Wen-qing

机构信息

College of Chemistry and Chemical Engineering, Jinggangshan College, Jiangxi Province Key Laboratory of Coordination Chemistry, Ji'an 343009, China.

出版信息

Guang Pu Xue Yu Guang Pu Fen Xi. 2007 Oct;27(10):2098-101.

Abstract

Small molecular water-miscible organic solvent, such as ethanol, acetone and acetonitrile, can be used to form aqueous two-phase system for extraction of analytes in the presence of salts. Just based on it, the acetone-salt-H2O and the ethanol-salt-H2O systems were developed for the determination of trace vitamin B2 coupled with fluorimetry after aqueous two-phase extraction. Various factors including the type and amount of salts and organic solvents, extraction time, coexistent substance, and pH were studied. The experimental results show that the extraction efficiencies of vitamin B2 in the acetone-salt-H2O and the ethanol-salt-H2O system were 97.6%-98.4% and 94.2%-95.5%, respectively, in one-step extraction. Under the optimal conditions, these two extraction techniques gave linear dynamic ranges of 1.5 x 10(-7)-3.5 x 10(-6) mol x L(-1) and 2..0 x 10(-7)-1.5 x 10(-6) mol x L(-1), with detection limits of 0.031 and 0.041 microg, respectively. Both techniques were applied directly to determination of vitamin B2 in pharmaceutical tablets and injections with satisfactory results. The relative errors were--1.8%-1.1% as compared with China pharmacopoeia and the relative standard deviations were 1.5%-2.2% for vitamin B2 in the above samples. Compared with conventional liquid-liquid extraction or solid-phase extraction, these methods have outstanding advantages such as simple operation, high extraction efficiency, little interference with analytes and low toxicity.

摘要

小分子与水互溶的有机溶剂,如乙醇、丙酮和乙腈,可用于在盐存在的情况下形成双水相系统以萃取分析物。基于此,开发了丙酮-盐-H₂O和乙醇-盐-H₂O系统,用于在双水相萃取后结合荧光法测定痕量维生素B₂。研究了各种因素,包括盐和有机溶剂的类型和用量、萃取时间、共存物质和pH值。实验结果表明,在一步萃取中,维生素B₂在丙酮-盐-H₂O和乙醇-盐-H₂O系统中的萃取效率分别为97.6%-98.4%和94.2%-95.5%。在最佳条件下,这两种萃取技术的线性动态范围分别为1.5×10⁻⁷-3.5×10⁻⁶ mol·L⁻¹和2.0×10⁻⁷-1.5×10⁻⁶ mol·L⁻¹,检测限分别为0.031和0.041 μg。这两种技术都直接应用于药物片剂和注射液中维生素B₂的测定,结果令人满意。与中国药典相比,相对误差为-1.8%-1.1%,上述样品中维生素B₂的相对标准偏差为1.5%-2.2%。与传统的液-液萃取或固相萃取相比,这些方法具有操作简单、萃取效率高、对分析物干扰小和毒性低等突出优点。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验