Cluley S R, Brener B J, Hollier L H, Eisenbud D E, Ferrara-Ryan M, Villaneuva A, Parsonnet V, Vilkomerson D
Division of Vascular Surgery, Newark Beth Israel Medical Centre, New Jersey 07112.
Am J Surg. 1991 Aug;162(2):117-21. doi: 10.1016/0002-9610(91)90171-9.
A new catheter has been developed that can be accurately and precisely positioned using duplex ultrasonography alone. A piezoelectric transducer that functions as a passive, omnidirectional receiver is attached to the mid-balloon region of an angioplasty catheter. Integration to a standard duplex imaging system allows visualization of this receiver in all planes, with the location of the receiver represented by a flashing bright arrow superimposed on the ultrasound B-mode image. This catheter was tested successfully in 10 dogs with bilateral femoral arteriovenous fistulas. With the use of duplex guidance, eight arterial and three venous stenoses were dilated and two stents were placed. Catheter placements and therapeutic outcomes predicted by ultrasound correlated with arteriographic findings. Ultrasound guidance has several advantages. It eliminates the risks associated with radiographic imaging. It also reduces patient costs and discomfort. Furthermore, procedures can be monitored continuously in real time with the collection of anatomic and physiologic data.