Suppr超能文献

A new technique for right heart catheterization with a Swan-Ganz balloon catheter via femoral vein in the patient with dilated right heart chambers, utilizing a modified transseptal-type sheath.

作者信息

Lee H H, Mahrer P R

机构信息

Section of Cardiology, Kaiser Permanente Harbor City Medical Center, CA 90710.

出版信息

Cathet Cardiovasc Diagn. 1991 May;23(1):54-6. doi: 10.1002/ccd.1810230115.

Abstract

Right heart catheterization with a Swan-Ganz balloon catheter via femoral vein in the patient with dilated right atrium and ventricle is frequently difficult due to excessive catheter coiling in the dilated atrium. A new technique is described in this situation. By using a modified transseptal-type sheath positioned counterclockwise in right atrium as an introducing sheath, a balloon catheter can be easily advanced to the distal pulmonary artery by smooth counterclockwise direction and secure support from this sheath.

摘要