Nakagawa Saori, Suzuki Riho, Yamazaki Reiko, Kusuhara Yoko, Mitsumoto Shoko, Kobayashi Hikaru, Shimoeda Sadahiko, Ohta Shin, Yamato Susumu
Department of Bio-analytical Chemistry, Faculty of Pharmaceutical Sciences, Niigata University of Pharmacy and Applied Life Sciences, 265-1 Higashijima, Akiha-ku, Niigata 956-8603, Japan.
Chem Pharm Bull (Tokyo). 2008 Mar;56(3):328-31. doi: 10.1248/cpb.56.328.
A simple column-switching high performance liquid chromatographic (HPLC) method that does not require any complicated pretreatment has been developed to determine voriconazole in human plasma samples. An internal standard (IS) and borate buffer (pH 9.0) were added to plasma samples, which were then injected directly into the column-switching HPLC system using MAYI-ODS as a pre-column. The calibration curve for voriconazole showed good linearity in the range of 0.2-10 mug/ml in human plasma. The mean RSD (%) value of intra-day (n=6) and inter-day (n=5) precision were less than 5.4% and 8.2%, respectively. This system could make more than three hundred successive, accurate measurements when a washing step with ammonium acetate solution was added. This method was successfully applied to measure the therapeutic voriconazole level in patients' plasma, and was used in a study of voriconazole pharmacokinetics after oral administration.
已开发出一种简单的柱切换高效液相色谱(HPLC)方法,该方法无需任何复杂的预处理即可测定人血浆样品中的伏立康唑。将内标(IS)和硼酸盐缓冲液(pH 9.0)添加到血浆样品中,然后使用MAYI-ODS作为预柱直接将其注入柱切换HPLC系统。伏立康唑的校准曲线在人血浆中0.2-10微克/毫升范围内显示出良好的线性。日内(n=6)和日间(n=5)精密度的平均相对标准偏差(RSD)(%)值分别小于5.4%和8.2%。当添加用醋酸铵溶液的洗涤步骤时,该系统可以进行三百多次连续、准确的测量。该方法已成功应用于测定患者血浆中的伏立康唑治疗水平,并用于口服给药后伏立康唑药代动力学的研究。