El-Saba A M, Azzam R M, Abushagur M A
Department of Electrical and Computer Engineering, University of Alabama at Huntsville, Huntsville, Alabama 35899, USA.
Appl Opt. 1999 May 1;38(13):2829-36. doi: 10.1364/ao.38.002829.
A division-of-amplitude photopolarimeter that uses a parallel-slab multiple-reflection beam splitter was described recently [Opt. Lett. 21, 1709 (1996)]. We provide a general analysis and an optimization of a specific design that uses a fused-silica slab that is uniformly coated with a transparent thin film of ZnS on the front surface and with an opaque Ag or Au reflecting layer on the back. Multiple internal reflections within the slab give rise to a set of parallel, equispaced, reflected beams numbered 0, 1, 2, and 3 that are intercepted by photodetectors D(0), D(1), D(2), and D(3), respectively, to produce output electrical signals i(0), i(1), i(2), and i(3), respectively. The instrument matrix A, which relates the output-signal vector I to the input Stokes vector S by I = AS, and its determinant D are analyzed. The instrument matrix A is nonsingular; hence all four Stokes parameters can be measured simultaneously over a broad spectral range (UV-VIS-IR). The optimum film thickness, the optimum angle of incidence, and the effect of light-beam deviation on the measured input Stokes parameters are considered.
最近描述了一种使用平行平板多次反射分束器的振幅分割式光偏振计[《光学快报》21, 1709 (1996)]。我们对一种特定设计进行了全面分析和优化,该设计使用的熔石英平板,其前表面均匀涂覆有一层透明的硫化锌薄膜,后表面涂覆有不透明的银或金反射层。平板内的多次内反射产生一组平行、等间距的反射光束,编号为0、1、2和3,分别被光电探测器D(0)、D(1)、D(2)和D(3)拦截,以分别产生输出电信号i(0)、i(1)、i(2)和i(3)。分析了通过I = AS将输出信号向量I与输入斯托克斯向量S相关联的仪器矩阵A及其行列式D。仪器矩阵A是非奇异的;因此,所有四个斯托克斯参数可以在很宽的光谱范围(紫外 - 可见 - 红外)内同时测量。考虑了最佳薄膜厚度、最佳入射角以及光束偏差对测量的输入斯托克斯参数的影响。