镍钛形状记忆合金舟骨弧形钉的研制与生物力学研究

[Development and biomechanical study of Ni-Ti shape memory alloys scaphoid arc nail].

作者信息

Kong Weiyun, Xu Yongqing, Wang Yufei, Chen Bin, Xu Zhongyong, Li Rui

机构信息

Orthopedic Center, Kunming General Hospital of Chengdu Military Command of PLA, Kunming Yunnan, 650032, P. R. China.

出版信息

Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2008 Jan;22(1):48-52.

Abstract

OBJECTIVE

To develop a new internal fixator Ni-Ti shape memory alloy scaphoid arc nail (NT-SAN) for fracture of the scaphoid waist and to test the biomechanical characteristics of NT-SAN so as to provide the biomechanical proofs for clinical application.

METHODS

According to the acquired measurements and anatomic features of the scaphoid in Chinese population, a special internal fixator--NT-SAN was designed. The biomechanical intensity experiment: The 32 specimens of fractures of the waist of scaphoid were divided randomly into 4 groups (n=8). Reduction and internal fixation were carried out in each specimen, with Kirschner wires (Group A), with Kirschner wires stapling (Group B), with screw (Group C) and with NT-SAN (Group D). Their fixed strength was tested by a hydrolicpressure biomechanical system AG-1. The biomechanical fatigue experiment: The models of 24 waist scaphoid fracture of adult upper limbs specimens were made and randomly divided into 3 groups (n=8). Fracture was fixed with Kirschner wire stapling (Group E), with screw(Group F) and with NT-SAN (Group G). Wrist joint was vertically pressured load of 98 N, palmar flexion and dorsal extension motion was simulated; the range of movement was from palmar flexion 5 degrees to dorsal extension 30 degrees and the frequency was 2,000 cycles. Displace data of fragment were measured at every 500 cycles. Finally, the biomechanical features of NT-SAN were evaluated according to the tested data. NT-SAN was used to treated 1 patient with fracture of the waist of scaphoid, who was typed as IIb according to Herbert classification.

RESULTS

The biomechanical tests showed that the traction forces were (15.18+/-3.55), (36.04+/-4.30), (64.88+/-11.62), (65.84+/-12.22) N and (20.28+/-12.09), (75.95+/-47.64), (120.91+/-26.68), (130.21+/-31.55) N when the displacements of the fracture lines distracted in 1 mm/min were 1 mm and 2 mm; showing significant differences between Group D and groups A, B (P < 0.05), and showing no significant difference between Group D and Group C (P < 0.05). The biomechanical fatigue experiment showed that there were significant differences between Group G and groups E, F (P < 0.05) according to the measured results of the displacements of the fracture lines. When wrist joint were circularly moved; showing significant differences between Group G and Group E (P < 0.05) after 500 circular movements, and showing significant differences between Group G and Group F (P < 0.05) after 1,500 circular movements according to the measured results of the "stage-shape" displacements in the fracture position. Incision healed by first intention and the X-ray films showed good NT-SAN fixation 3 months after follow-up.

CONCLUSION

The design of NT-SAN is in accordance with the anatomic features of the scaphoid. The fixed strength can meet the demand of the biomechanics and the range of fatigue strength can meet the demand of bony union.

摘要

目的

研制一种新型内固定器镍钛形状记忆合金舟骨弧形钉(NT-SAN)用于舟骨腰部骨折,测试NT-SAN的生物力学特性,为临床应用提供生物力学依据。

方法

根据国人舟骨的测量数据及解剖特点,设计一种特殊的内固定器——NT-SAN。生物力学强度实验:将32例舟骨腰部骨折标本随机分为4组(n = 8)。每组标本均行复位内固定,A组用克氏针,B组用克氏针环扎,C组用螺钉,D组用NT-SAN。用液压生物力学系统AG-1测试其固定强度。生物力学疲劳实验:制作24例成人上肢舟骨腰部骨折模型,随机分为3组(n = 8)。E组用克氏针环扎固定骨折,F组用螺钉固定,G组用NT-SAN固定。对腕关节施加垂直压力98 N,模拟掌屈和背伸运动;运动范围为掌屈5度至背伸30度,频率为2 000次循环。每500次循环测量骨折块移位数据。最后根据测试数据评估NT-SAN的生物力学特性。应用NT-SAN治疗1例Herbert分型为IIb型的舟骨腰部骨折患者。

结果

生物力学测试显示,当骨折线以1 mm/min的速度牵开1 mm和2 mm时,牵拉力分别为(15.18±3.55)、(36.04±4.30)、(64.88±11.62)、(65.84±12.22)N和(20.28±12.09)、(75.95±47.64)、(120.91±26.68)、(130.21±31.5)N;D组与A、B组比较差异有统计学意义(P < 0.05),D组与C组比较差异无统计学意义(P > 0.05)。生物力学疲劳实验显示,根据骨折线移位测量结果,G组与E、F组比较差异有统计学意义(P < 0.05)。当腕关节做环形运动时,根据骨折部位“阶梯状”移位测量结果,G组与E组在500次环形运动后差异有统计学意义(P < 0.05),G组与F组在1 500次环形运动后差异有统计学意义(P < 可见手术切口一期愈合,随访3个月X线片显示NT-SAN固定良好。

结论

NT-SAN的设计符合舟骨解剖特点,固定强度能满足生物力学要求,疲劳强度范围能满足骨愈合要求。 0.05)。