Suppr超能文献

通过在线固相萃取-液相色谱-电喷雾串联质谱法对污水中纳克每升水平的不同种类滥用药物进行全自动测定。

Fully automated determination in the low nanogram per liter level of different classes of drugs of abuse in sewage water by on-line solid-phase extraction-liquid chromatography-electrospray-tandem mass spectrometry.

作者信息

Postigo Cristina, de Alda Maria J Lopez, Barceló Damià

机构信息

Department of Environmental Chemistry, IIQAB-CSIC, c/Jordi Girona 18-26, 08034 Barcelona, Spain.

出版信息

Anal Chem. 2008 May 1;80(9):3123-34. doi: 10.1021/ac702060j. Epub 2008 Apr 1.

Abstract

The present work describes the first fully automated method, based on on-line solid-phase extraction (SPE)-liquid chromatography-electrospray-tandem mass spectrometry, developed for the determination of drugs of abuse (17 compounds and metabolites belonging to the classes of amphetaminics, cannabinoids, cocainics, opiates, and lysergics) in sewage waters. On-line SPE is performed by passing 5 mL of the water sample through a PLRP-s cartridge for analytes measured in positive ionization mode (all but cannabinoids) and through an Oasis HLB cartridge for analytes measured in negative ionization mode (cannabinoids). For unequivocal identification and confirmation two selected reaction monitoring transitions are registered per compound, thus achieving the four identification points requested by the European Union for banned substances. Quantitation is performed by the internal standard method, indispensable to correct for matrix effects. The main advantages of the method developed are high sensitivity (limits of determination between 0.69 and 5.97 ng/L), selectivity and reliability of results, minimum sample manipulation, full automation, and fairly high throughput (analysis time per sample is 2 x 35 min). As a part of the validation procedure, the method developed has been applied to the analysis of various influent and effluent samples from four Spanish sewage treatment plants.

摘要

本研究描述了首个基于在线固相萃取(SPE)-液相色谱-电喷雾串联质谱法的全自动方法,该方法用于测定污水中的滥用药物(17种化合物及其代谢物,属于苯丙胺类、大麻素类、可卡因类、阿片类和麦角酸类)。在线固相萃取通过使5 mL水样流经PLRP-s柱(用于以正离子模式测定的分析物,大麻素类除外)和Oasis HLB柱(用于以负离子模式测定的分析物,即大麻素类)来进行。为实现明确鉴定和确认,每种化合物记录两个选择反应监测跃迁,从而达到欧盟对违禁物质要求的四个鉴定点。定量采用内标法,这对于校正基质效应必不可少。所开发方法的主要优点包括高灵敏度(测定限在0.69至5.97 ng/L之间)、结果的选择性和可靠性、最少的样品处理、全自动操作以及相当高的通量(每个样品的分析时间为2×35分钟)。作为验证程序的一部分,所开发的方法已应用于对来自四个西班牙污水处理厂的各种进水和出水样品的分析。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验