A potential danger of flexible endoscopy sheaths: a detached tip and how to retrieve it.

作者信息

Awad Z, Pothier D D

机构信息

Department of Otolaryngology, Head and Neck Surgery, St Bartholomew's Hospital, London, UK.

出版信息

J Laryngol Otol. 2009 Feb;123(2):243-4. doi: 10.1017/S0022215108002107. Epub 2008 Apr 3.

Abstract

OBJECTIVE

To report an unrecognised complication of fibre-optic nasendoscopy, and its management.

CASE REPORT

A protective, transparent nasendoscopy sheath is often used to reduce nasendoscope 'downtime' and to prevent cross infection, with minimal effect on the obtained image quality. We report the case of a subcutaneous tracheostomy procedure during which, without undue strain, the tip of the sheath became detached and acted as a foreign body within the trachea. A urological stone retrieval basket was used to retrieve the sheath, after failure of conventional methods.

DISCUSSION

Clinicians should be aware that any instrument introduced into the airway has the potential to fail and in the process produce a foreign body which may cause serious complications. The urological stone retrieval basket may be a useful addition to the current set of instruments used to deal with difficult airway foreign bodies.