Liu Chia-Ying, Varadarajan Padmini, Pohost Gerald M, Nayak Krishna S
Viterbi School of Engineering, University of Southern California, Los Angeles, CA, USA.
Magn Reson Imaging. 2008 Jun;26(5):661-6. doi: 10.1016/j.mri.2008.01.001.
The purpose of this study was to demonstrate and evaluate the performance of real-time color-flow MRI at 3 T using variable-density spiral (VDS) phase contrast. Spiral phase contrast imaging was implemented within a flexible real-time interactive MRI system that provided continuous image reconstruction and an intuitive user interface. The pulse sequence consisted of a spectral-spatial excitation, bipolar gradient, spiral readout and gradient spoiler. VDSs were utilized to increase spatial and/or temporal resolution relative to uniform-density spirals (UDSs). Parameter choices were guided by specific applications. Sliding window reconstruction was used to achieve a maximum display rate of 40 frames/s. No breath-holding or gating was used. Our results demonstrated that real-time color-flow movies using UDS and VDS provided adequate visualization of intracardiac flow, carotid flow and proximal coronary flow in healthy volunteers. Average aortic outflow velocity measured at the aortic valve plane using VDS was 29.4% higher than that using UDS. Peak velocity measured in the common carotid artery using VDS was 9.8% higher than that using UDS.
本研究的目的是演示和评估使用可变密度螺旋(VDS)相位对比的3T实时彩色血流磁共振成像(MRI)的性能。螺旋相位对比成像在一个灵活的实时交互式MRI系统中实现,该系统提供连续图像重建和直观的用户界面。脉冲序列由频谱空间激发、双极梯度、螺旋读出和梯度扰相器组成。相对于均匀密度螺旋(UDS),VDS用于提高空间和/或时间分辨率。参数选择由特定应用指导。使用滑动窗口重建以实现40帧/秒的最大显示速率。未使用屏气或门控。我们的结果表明,使用UDS和VDS的实时彩色血流电影能够充分显示健康志愿者的心内血流、颈动脉血流和近端冠状动脉血流。使用VDS在主动脉瓣平面测得的平均主动脉流出速度比使用UDS高29.4%。使用VDS在颈总动脉测得的峰值速度比使用UDS高9.8%。