Yang Huili, Aboyewa Oluyemi B, Webster Gregory, Shah Dhaivat, Golestanirad Laleh, Baraboo Justin J, Markl Michael, Collins Jeremy D, Knight Bradley P, Hong KyungPyo, Patel Amit R, Lee Daniel C, Kim Daniel
Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, Illinois, USA.
Magn Reson Med. 2025 May;93(5):2099-2107. doi: 10.1002/mrm.30399. Epub 2024 Dec 6.
Overall there is a lack of evidence on the accuracy and precision of phase-contrast (PC) MRI in patients with cardiac implantable electronic devices (CIEDs). The purpose of this study is to determine whether aortic velocity measurements are influenced by a CIED.
We scanned 21 adult patients and 8 pediatric volunteers using clinical standard PC and real-time PC (rt-PC) sequences with and without a CIED generator taped on human subjects (below the left clavicle for adults and children; also on the abdomen for children) to mimic image artifacts. Peak and mean velocities above the aortic valve were calculated.
The Bland-Altman analyses on peak velocity measurements in pediatric subjects showed that both the accuracy and precision worsen as the distance between the CIED and aortic valve decreases (i.e., from abdomen to below the left clavicle). Specifically, both the bias and the coefficient of variation (CV) for both clinical PC and rt-PC increased from the abdominal position (clinical: bias = -1.1%, CV = 4.3%; rt-PC: bias = -0.3%, CV = 3.4%) to the clavicle position (clinical: bias = -4.0%, CV = 8.1%; rt-PC: bias = 8.2%, CV = 7.3%). A similar trend was observed for mean velocity. The mean difference in peak and mean velocity measurements between rt-PC with CIED (either position) and clinical standard PC with no CIED was within 7.5%. In adult patients, the mean difference between rt-PC with CIED and clinical standard PC with CIED in peak velocity was 6.9%, and the CV was 7.9%.
This study demonstrates that aortic velocity measurements are influenced by CIED in both adult and pediatric subjects.
总体而言,缺乏关于心脏植入式电子设备(CIED)患者相位对比(PC)MRI准确性和精确性的证据。本研究的目的是确定主动脉速度测量是否受CIED影响。
我们使用临床标准PC序列和实时PC(rt-PC)序列对21名成年患者和8名儿科志愿者进行扫描,在人体受试者(成人和儿童的左锁骨下方;儿童还包括腹部)粘贴和不粘贴CIED发生器以模拟图像伪影。计算主动脉瓣上方的峰值速度和平均速度。
对儿科受试者峰值速度测量的Bland-Altman分析表明,随着CIED与主动脉瓣之间距离的减小(即从腹部到左锁骨下方),准确性和精确性均变差。具体而言,临床PC和rt-PC的偏差和变异系数(CV)均从腹部位置(临床:偏差=-1.1%,CV=4.3%;rt-PC:偏差=-0.3%,CV=3.4%)增加到锁骨位置(临床:偏差=-4.0%,CV=8.1%;rt-PC:偏差=8.2%,CV=7.3%)。平均速度也观察到类似趋势。有CIED(任一位置)的rt-PC与无CIED的临床标准PC之间的峰值和平均速度测量的平均差异在7.5%以内。在成年患者中,有CIED的rt-PC与有CIED的临床标准PC在峰值速度上的平均差异为6.9%,CV为7.9%。
本研究表明,成年和儿科受试者的主动脉速度测量均受CIED影响。