Lehmann Kai S, Ritz Joerg-P, Valdeig Steffi, Schenk Andrea, Holmer Christoph, Peitgen Heinz-O, Buhr Heinz-J, Frericks Bernd B
Department of General, Vascular and Thoracic Surgery-Chirurgische Klinik I, Charité-Campus Benjamin Franklin, Freie- und Humboldt-Universität zu Berlin, Hindenburgdamm 30, 12200, Berlin, Germany.
Ann Surg Oncol. 2008 Jul;15(7):1899-907. doi: 10.1245/s10434-008-9934-x. Epub 2008 May 1.
Computer systems allow the planning of complex liver operations. The segmentation of intrahepatic vessels builds the basis for the calculation of liver segments and resection proposals. For surgical use, it is essential to know the capabilities and limitations of the segmentation. The aim of this study was to determine the sensitivity and precision of the portal vein segmentation of a computer planning system for liver surgery in vivo.
Segmentations were performed with the software system HepaVision on computed tomography (CT) scan data of domestic pigs. An in situ corrosion cast of the portal vein served as the gold standard. The segmentation results of the portal vein and the corrosion cast were compared with regard to sensitivity, precision, and amount of short-circuit segmentations.
The methodology demonstrated high resolution ex situ. The in vivo sensitivity of the portal vein segmentation was 100% for vessels of more than 5 mm in diameter and 82% for vessels of 3-4 mm. All segment branches were detected as well as 84% of the first subsegment branches with a diameter of more than 3 mm. The precision of the system was 100% for segment branches and 89% for the first subsegment vessels. The amount of internal short-circuit segmentations was less than 3.0%. No external short-circuits were found.
The system has a high precision and sensitivity under clinical conditions. The segmentation is suitable for portal vein branches of the first and second order and for vessels of >/=3 mm in diameter.
计算机系统可用于复杂肝脏手术的规划。肝内血管的分割是肝段计算和切除方案制定的基础。对于手术应用而言,了解分割的能力和局限性至关重要。本研究的目的是确定一种用于肝脏手术的计算机规划系统在体内对门静脉分割的敏感性和精确性。
使用HepaVision软件系统对家猪的计算机断层扫描(CT)数据进行分割。门静脉的原位腐蚀铸型作为金标准。将门静脉的分割结果与腐蚀铸型在敏感性、精确性和短路分割量方面进行比较。
该方法在体外显示出高分辨率。门静脉分割在体内对直径大于5mm的血管敏感性为100%,对直径3 - 4mm的血管敏感性为82%。所有段分支以及84%直径大于3mm的一级亚段分支均被检测到。该系统对段分支的精确性为100%,对一级亚段血管的精确性为89%。内部短路分割量小于3.0%。未发现外部短路。
该系统在临床条件下具有较高的精确性和敏感性。该分割适用于一级和二级门静脉分支以及直径≥3mm的血管。