Two confounding factors were selected and analyzed in blood pressure estimation using pulse arrival time (PAT) for each individual. The heart rate was used as the confounding factor for the cardiac cycle, and the duration from the maximum derivative point to the dicrotic peak (TDB) in the photoplethysmogram was used as another confounding factor representing arterial stiffness. By considering these factors with PAT in multiple regression analysis, the performance of blood pressure estimation is enhanced significantly in the diastolic phase as well as in the systolic phase. The reproducibility of this method was also validated with formerly obtained regression equations from the training set. The correlation between estimated and measured blood pressure decreased a little, but the validity was still maintained (r congruent with 0.8). This shows the value of the method in non-intrusive blood pressure estimation for individual patients and may be useful for various applications.
在使用脉搏波传导时间(PAT)对每个个体进行血压估计时,选择并分析了两个混杂因素。心率被用作心动周期的混杂因素,而光电容积脉搏波描记图中从最大导数点到重搏波峰的持续时间(TDB)被用作另一个代表动脉僵硬度的混杂因素。通过在多元回归分析中将这些因素与PAT相结合,在舒张期和收缩期,血压估计的性能都得到了显著提高。该方法的可重复性也通过先前从训练集中获得的回归方程得到了验证。估计血压与测量血压之间的相关性略有下降,但有效性仍然保持(r约为0.8)。这表明该方法在个体患者的无创血压估计中的价值,并且可能对各种应用有用。