Suppr超能文献

Development of a highly precise ID-ICP-SFMS method for analysis of low concentrations of lead in rice flour reference materials.

作者信息

Zhu Yanbei, Inagaki Kazumi, Yarita Takashi, Chiba Koichi

机构信息

National Metrology Institute of Japan, 1-1-1, Umezono, Tsukuba, Ibaraki, 305-8563, Japan.

出版信息

Anal Bioanal Chem. 2008 Jul;391(6):2055-60. doi: 10.1007/s00216-008-2153-2. Epub 2008 May 27.

Abstract

Microwave digestion and isotope dilution inductively coupled plasma mass spectrometry (ID-ICP-SFMS) has been applied to the determination of Pb in rice flour. In order to achieve highly precise determination of low concentrations of Pb, the digestion blank for Pb was reduced to 0.21 ng g(-1) after optimization of the digestion conditions, in which 20 mL analysis solution was obtained after digestion of 0.5 g rice flour. The observed value of Pb in a non-fat milk powder certified reference material (CRM), NIST SRM 1549, was 16.8 +/- 0.8 ng g(-1) (mean +/- expanded uncertainty, k = 2; n = 5), which agreed with the certified value of 19 +/- 3 ng g(-1) and indicated the effectiveness of the method. Analytical results for Pb in three brown rice flour CRMs, NIST SRM 1568a, NIES CRM 10-a, and NIES CRM 10-b, were 7.32 +/- 0.24 ng g(-1) (n = 5), 1010 +/- 10 ng g(-1) (n = 5), and 1250 +/- 20 ng g(-1) (n = 5), respectively. The concentration of Pb in a candidate white rice flour reference material (RM) sample prepared by the National Metrology Institute of Japan (NMIJ) was observed to be 4.36 +/- 0.28 ng g(-1) (n = 10 bottles).