Munk-Hartig Anne-Kathrin, Stübinger Sven Henrik, Jünemann Klaus-Peter, van der Horst Christof
Department of Urology, University Clinic Schleswig-Holstein, Campus Kiel, Kiel, Germany.
J Sex Med. 2008 Jul;5(7):1758-61. doi: 10.1111/j.1743-6109.2008.00849.x.
The renewal of penile prosthesis is a foreseeable intervention because of the limited survival time of the different implants. It usually becomes necessary simply for material fatigue after long-term regular use; in some cases, premature intervention might be necessary because of infection or other perioperative complications.
In the case presented here, the removal of a Dynaflex penile prosthesis showed a missing rear-tip extender on the right-hand tube.
A cavernoscopy of the right cavernous body was performed by using a conventional urethrocystoscope. The found rear-tip extender was extracted by using alligator forceps via the cystoscopic device.
A new AMS CX hydraulic cavernous body prosthesis was successfully implanted. It has proven fully functional up to the current follow-up control.
This case shows that cavernoscopy is a gentle intraoperative procedure for diagnosing and removing material in penile prosthesis surgery.
由于不同阴茎假体的使用寿命有限,阴茎假体的更换是一种可预见的干预措施。长期正常使用后,通常仅因材料疲劳就有必要进行更换;在某些情况下,由于感染或其他围手术期并发症,可能需要提前干预。
在此呈现的病例中,取出一个Dynaflex阴茎假体时发现右侧管道上的后端延长器缺失。
使用传统尿道膀胱镜对右侧海绵体进行海绵体镜检查。通过膀胱镜设备用鳄鱼钳取出找到的后端延长器。
成功植入了一个新的AMS CX液压海绵体假体。截至目前的随访检查,其功能已得到充分验证。
该病例表明,海绵体镜检查是阴茎假体手术中用于诊断和取出材料的一种温和的术中操作。