Retinal nerve fiber layer measurement repeatability in scanning laser polarimetry with enhanced corneal compensation.

作者信息

Mai Thê Anh, Reus Nicolaas J, Lemij Hans G

机构信息

The Rotterdam Eye Hospital, Rotterdam, The Netherlands.

出版信息

J Glaucoma. 2008 Jun-Jul;17(4):269-74. doi: 10.1097/IJG.0b013e31815c3a6b.

Abstract

OBJECTIVE

To investigate the measurement repeatability of the various standard retinal nerve fiber layer (RNFL) parameters in scanning laser polarimetry (SLP) with enhanced corneal compensation (ECC, released for investigational use on the commercially available GDx with variable corneal compensation, Carl Zeiss Meditec Inc, Dublin, CA) in healthy eyes, eyes with ocular hypertension (OHT), and glaucomatous eyes.

DESIGN

Cross-sectional observational study.

PARTICIPANTS

Sixteen healthy subjects, 32 subjects with OHT, and 35 patients with primary open-angle glaucoma.

METHODS

Eighty-three eyes of eighty-three white participants were imaged with SLP ECC 3 times on a same day. We determined the intraeye RNFL measurement repeatability.

MAIN OUTCOME MEASURES

Intraeye within-subject standard deviation (Sw), repeatability coefficient, and the 2-way mixed intraclass correlation coefficient for various standard RNFL parameters in SLP ECC.

RESULTS

The repeatability of the RNFL measurements was generally good. In glaucomatous eyes, the Sw and repeatability coefficient for the nerve fiber indicator and temporal-superior-nasal-inferior-temporal average were statistically significantly higher than in healthy eyes and eyes with OHT. In each group, the intraclass correlation coefficient values were generally similar across the parameters, except for the intereye symmetry. The Sw values for various parameters were generally considerably less than 9% of the measurement spectrum.

CONCLUSIONS

RNFL measurements by SLP ECC had, in general, a good measurement repeatability, although some parameters seemed to be less stable in glaucomatous eyes than in healthy eyes and eyes with OHT. SLP ECC may therefore probably be employed for the detection of glaucomatous progression.

摘要

目的

研究在使用增强角膜补偿(ECC,在卡尔蔡司医疗技术公司位于加利福尼亚州都柏林市的可变角膜补偿商用GDx上供研究使用)的扫描激光偏振仪(SLP)中,健康眼、高眼压症(OHT)眼和青光眼眼中各种标准视网膜神经纤维层(RNFL)参数的测量重复性。

设计

横断面观察性研究。

参与者

16名健康受试者、32名高眼压症受试者和35名原发性开角型青光眼患者。

方法

83名白人参与者的83只眼睛在同一天用SLP ECC进行3次成像。我们测定了眼内RNFL测量的重复性。

主要观察指标

SLP ECC中各种标准RNFL参数的眼内受试者内标准差(Sw)、重复性系数和双向混合组内相关系数。

结果

RNFL测量的重复性总体良好。在青光眼眼中,神经纤维指标以及颞侧-上方-鼻侧-下方-颞侧平均值的Sw和重复性系数在统计学上显著高于健康眼和高眼压症眼。在每组中,除了双眼对称性外,各参数的组内相关系数值总体相似。各种参数的Sw值总体上远小于测量范围的9%。

结论

一般而言,SLP ECC进行的RNFL测量具有良好的测量重复性,尽管某些参数在青光眼眼中似乎比在健康眼和高眼压症眼中更不稳定。因此,SLP ECC可能可用于检测青光眼的进展。