Hawker Matthew J, Ainsworth Gerard, Vernon Stephen A, Dua Harminder S
Department of Ophthalmology, Queen's Medical Centre, Derby Road, Nottingham, NG7 2UH, UK.
J Glaucoma. 2008 Jun-Jul;17(4):280-6. doi: 10.1097/IJG.0b013e31815c3a33.
This study assessed agreement in optic nerve head parameters between 2 investigators independently contouring Heidelberg retina tomograph II (HRT II) optic disc images of normal elderly subjects.
Contours were drawn by both investigators on 550 randomly selected optic discs of 550 normal elderly patients (mean age, 73.1 y) drawn from a population-based study. All subjects had normal intraocular pressures and visual fields. Coefficients of variation were calculated for interobserver and intraobserver variability.
The mean disc area was 1.97 mm, with no relationship between magnitude of disagreement and the mean. There was a significant bias between investigators with a mean difference in disc area of -0.06 mm. The 95% limits of agreement of disc area were -0.59 to +0.47 mm. Ninety-five percent coefficients of agreement (1.96 SD of the parameter differences between investigators/mean parameter value x100) were similar for global and sectoral disc areas (26.7% to 32.3%). Respective coefficients for rim area, cup area, and rim volume were much higher at 35.2%, 48.5%, and 76.5% respectively, but much lower for rim/disc area ratio (12.7%). There was no significant correlation between disagreement and grade of cataract or HRT image quality.
Disagreement in contour placement can be substantial, with disproportionate effects on some calculated HRT parameters. Image quality and lens opacity have no significant influence on the degree of disagreement. Rim/disc area ratio is the HRT parameter least affected and may be more useful in comparing normative databases and diagnostic studies between centers.
本研究评估了两位研究者在独立勾勒正常老年受试者的海德堡视网膜断层扫描II(HRT II)视盘图像的视神经乳头参数时的一致性。
两位研究者从一项基于人群的研究中随机选取550名正常老年患者(平均年龄73.1岁)的550个视盘进行轮廓勾勒。所有受试者眼压和视野均正常。计算观察者间和观察者内变异性的变异系数。
平均视盘面积为1.97平方毫米,差异大小与平均值之间无相关性。研究者之间存在显著偏差,视盘面积平均差异为-0.06平方毫米。视盘面积的95%一致性界限为-0.59至+0.47平方毫米。整体和扇形视盘面积的95%一致性系数(研究者之间参数差异的1.96倍标准差/平均参数值×100)相似(26.7%至32.3%)。边缘面积、杯状面积和边缘体积的相应系数分别高得多,为35.2%、48.5%和76.5%,但边缘/视盘面积比的系数则低得多(12.7%)。差异与白内障分级或HRT图像质量之间无显著相关性。
轮廓放置的差异可能很大,对一些计算出的HRT参数有不成比例的影响。图像质量和晶状体混浊对视盘轮廓差异程度无显著影响。边缘/视盘面积比是受影响最小的HRT参数,在比较不同中心的标准数据库和诊断研究时可能更有用。