Calicchia A, Mazzei F, Dobici F, Paganini Fioratti M, Indovina P L
Laboratorio di Fisica-Istituto Superiore di Sanità-Roma.
Radiol Med. 1991 Jun;81(6):910-7.
The aim of this paper is the critical analysis of the results obtained with the quality control program in X-ray diagnosis called NEXT (Nationwide Evaluation of X-ray Trends). It is also aimed at comparing the above results with guidelines and trends of international radiology, for both technical features and patient protection from radiations. The main purpose of patient protection is the reduction of unnecessary exposure through the optimization of radiodiagnostic procedures. In order to identify the fields requiring the application of specific quality control programs, the Istituto Superiore di Sanità, together with ENEA/DISP have been running the NEXT program in Italy since 1977. The program, through data collection and measurements in more than 2,500 radiologic units, allowed both state of the art and progress in radiodiagnostic practice to be studied over a long period of time. Data computation provides, for a standard patient, the skin entrance exposure and the main organ doses for the examinations which are most frequently performed. In this paper, the fundamentals of the NEXT program are reported and the most significant results are discussed, and compared, when possible, with the values recommended by international scientific organizations. The analysis of the results demonstrates a decreasing trend in patient exposure over the last years.
本文旨在对名为NEXT(全国X射线趋势评估)的X射线诊断质量控制项目所取得的结果进行批判性分析。同时,本文还旨在将上述结果与国际放射学的指南和趋势进行比较,包括技术特征和患者辐射防护方面。患者防护的主要目的是通过优化放射诊断程序来减少不必要的辐射暴露。为了确定需要应用特定质量控制项目的领域,自1977年以来,意大利高等卫生研究院与国家新技术、能源和可持续经济发展署(ENEA)/DISP一直在意大利开展NEXT项目。该项目通过在2500多个放射科单位收集数据和进行测量,使得长期研究放射诊断实践的现状和进展成为可能。数据计算为标准患者提供了最常进行的检查的皮肤入口处辐射剂量和主要器官剂量。本文报告了NEXT项目的基本情况,并讨论了最显著的结果,并在可能的情况下与国际科学组织推荐的值进行了比较。结果分析表明,在过去几年中患者的辐射暴露呈下降趋势。